FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1011843
·
Received March 11, 2008
Report
- Report Number
- 1720753-2008-17893
- Event Type
- Malfunction
- Date Received
- March 11, 2008
- Date of Event
- February 29, 2008
- Report Date
- March 7, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. REPLACED THE X-RAY TUBE. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SVC.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYS IS HAVING HIGH VOLTAGE ARCS. IT WAS NOTED THAT WHEN THE SYS TRACKS ABOVE 90KVP AN ARCING SOUND COMES FROM THE TUBE AND AN OVERLOAD FAULT MESSAGE IS PRESENTED. THERE WAS NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC | 9800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |