FDA Adverse Event Malfunction Summary report: N

9600

MDR report key: 1011838 · Received March 11, 2008

Report

Report Number
1720753-2008-17888
Event Type
Malfunction
Date Received
March 11, 2008
Date of Event
February 29, 2008
Report Date
March 7, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. ADJUSTED THE CAMERA FOCUS AND KV TRACKING. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SVC.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING QA TESTING OF THE 9600 SYS THE IMAGE QUAL WAS NOT AS GOOD AS EXPECTED. THERE WAS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC 9600 NA

Patients

Seq Age Sex Outcome Treatment
1