FDA Adverse Event Malfunction Summary report: N

AUTCAT2 WAVE

MDR report key: 1011618 · Received March 10, 2008

Report

Report Number
1219856-2008-00120
Event Type
Malfunction
Date Received
March 10, 2008
Date of Event
February 27, 2008
Report Date
March 10, 2008
Manufacturer
ARROW INTL., INC.
Product Code
DSP
PMA / PMN Number
K060309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

F/U REPORT WILL BE FILED IF ADD'L INFO BECOMES AVAILABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PT WAS AT THE HOSP WHEN THE INCIDENT OCCURRED. THE PUMP WAS EXCHANGED WITH ANOTHER PUMP THAT THE HOSP HAD. THERE WERE NO REPORTED PT COMPLICATIONS. THE FIELD SVC REPORT STATES THE FOLLOWING: SYMPTOM: UNIT STOPPED PUMPING WHILE ON PT. UNIT STOPPED PUMPING WHEN AC WAS REMOVED, UNIT NEVER SHUT OFF. FINDINGS/ACTIONS TAKEN: THE FIELD SVC REP WAS UNABLE TO DUPLICATE THE CUSTOMER'S COMPLAINT. UPON INITIAL INSPECTION, FOUND NUTS HOLDING KEY PAD TO DISPLAY HEAD OFF. AS A RESULT, THE KEY PAD WAS REATTACHED. FOUND THE MOTOR CABLE ATTACHED TO THE LEFT SIDE PANEL WITH CLAMP. THE CABLE WAS REATTACHED TO THE CHASSIS. ALL BUSHINGS MISSING ON BOTH SIDE PANELS. INSTALLED BUSHING. REPLACED WORN DISPLAY CABLE, REPLACED MISSING AC RETAINING CLAMP. THE UNIT PASSED ARROW FIELD SVC FUNCTIONAL CHECKOUT. RESULTS: EQUIPMENT OPERATIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTCAT2 WAVE INTRA-AORTIC BALLOON PUMP PRODUCTS DSP ARROW INTL., INC. NA

Patients

Seq Age Sex Outcome Treatment
1 UNK INTRA-AORTIC BALLOON PUMP (BRAND UNK)