FDA Adverse Event Malfunction Summary report: N

HT70 VENTILATOR

MDR report key: 10112298 · Received June 2, 2020

Report

Report Number
8020893-2020-00085
Event Type
Malfunction
Date Received
June 2, 2020
Date of Event
April 6, 2020
Report Date
June 2, 2020
Manufacturer
COVIDIEN
Product Code
CBK
UDI-DI
10884521544413
PMA / PMN Number
K111146
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: THE HT70 VENTILATOR WAS RETURNED TO MEDTRONIC FOR ANALYSIS. VISUAL INSPECTION OF THE BOARD REVEALED A DAMAGED U1104 CHIP OF THE SBC PCB. THE FAULT WAS ISOLATED TO THE SINGLE BOARD COMPUTER (SBC) PRINTED CIRCUIT BOARD (PCB). A NEW SBC PCB WAS REPLACED TO RESOLVE THE REPORTED ISSUE. THE VENTILATOR PASSED ALL TESTS AND CALIBRATIONS AS PER MANUFACTURER SPECIFICATIONS. THE LIKELY CAUSE OF THE EVENT WAS ISOLATED TO DAMAGED U1104 CHIP IN SBC BOARD. THE EVENT WILL BE INCLUDED IN TRENDING AND MONITORING. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING SERVICING, THE HT70 VENTILATOR SCREEN WAS OBSERVED TO BE FROZEN AND THE UNIT WAS LOCKED UP/ STOPPED VENTILATING. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME OF THE REPORTED EVENT.