BD PLASTIPAK LUER-LOK SYRINGE
Report
- Report Number
- 3003152976-2020-00235
- Event Type
- Malfunction
- Date Received
- June 2, 2020
- Date of Event
- May 13, 2020
- Report Date
- June 29, 2020
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- PHARMACIST
Narratives
H.6. INVESTIGATION: ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PHOTO WAS VISUALLY INSPECTED AND NO SIGNS OF EXCESS SILICONE CAN BE OBSERVED INSIDE OR OUTSIDE OF THE BARREL. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 1912280, NO DEVIATIONS OR NON-CONFORMANCES RELATED TO THIS ISSUE WERE IDENTIFIED DURING THE MANUFACTURING PROCESS . LUBRICANT IS EMPLOYED DURING THE SYRINGE ASSEMBLY PROCESS TO LUBRICATE THE CYLINDERS IN THE SILICONE STATION. THE SILICONE EMPLOYED IN THIS PRODUCT IS A MEDICAL GRADE SILICONE AUTHORIZED FOR PRODUCT USE. SILICONE CONTENT TESTS ARE PERFORMED DURING THE MANUFACTURING PROCESS OF EACH LOT NUMBER ACCORDING TO PROCEDURE, RESULTS WERE REVIEWED FOR THE REPORTED LOT AND FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME.
IT WAS REPORTED THAT A "GREASY" LAYER OF FOREIGN MATTER WAS VISIBLE INSIDE THE BD PLASTIPAK¿ LUER-LOK¿ SYRINGE BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM DUTCH TO ENGLISH: WE HAVE BEEN USING THIS SYRINGE FOR A LONG TIME WITHOUT PROBLEMS, WITH ONE OF THE CYTOSTATIC FILLERS, MITOMYCIN, A GREASY LAYER (SILICONE?) IS VISIBLE ON THE INSIDE OF THE SYRINGE. THIS IS REALLY NOT ON THE OUTSIDE, NOT TO WIPE OR SCRAPE OFF. I CANNOT SEND THIS SYRINGE BECAUSE OF THE RISKY CONTENT. WE TOOK THE REST OF THIS BATCH OF SYRINGES (12 PIECES) FROM THE PHARMACY SHELVES AND THEN USED ANOTHER LOT NUMBER WITHOUT ANY PROBLEMS.
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT A "GREASY" LAYER OF FOREIGN MATTER WAS VISIBLE INSIDE THE BD PLASTIPAK¿ LUER-LOK¿ SYRINGE BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: WE HAVE BEEN USING THIS SYRINGE FOR A LONG TIME WITHOUT PROBLEMS, WITH ONE OF THE CYTOSTATIC FILLERS, MITOMYCIN, A GREASY LAYER (SILICONE?) IS VISIBLE ON THE INSIDE OF THE SYRINGE. THIS IS REALLY NOT ON THE OUTSIDE, NOT TO WIPE OR SCRAPE OFF. I CANNOT SEND THIS SYRINGE BECAUSE OF THE RISKY CONTENT. WE TOOK THE REST OF THIS BATCH OF SYRINGES (12 PIECES) FROM THE PHARMACY SHELVES AND THEN USED ANOTHER LOT NUMBER WITHOUT ANY PROBLEMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571314 | BD PLASTIPAK LUER-LOK SYRINGE | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 1912280 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |