FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK LUER-LOK SYRINGE

MDR report key: 10111441 · Received June 2, 2020

Report

Report Number
3003152976-2020-00235
Event Type
Malfunction
Date Received
June 2, 2020
Date of Event
May 13, 2020
Report Date
June 29, 2020
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE PHOTO WAS PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THE PHOTO WAS VISUALLY INSPECTED AND NO SIGNS OF EXCESS SILICONE CAN BE OBSERVED INSIDE OR OUTSIDE OF THE BARREL. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 1912280, NO DEVIATIONS OR NON-CONFORMANCES RELATED TO THIS ISSUE WERE IDENTIFIED DURING THE MANUFACTURING PROCESS . LUBRICANT IS EMPLOYED DURING THE SYRINGE ASSEMBLY PROCESS TO LUBRICATE THE CYLINDERS IN THE SILICONE STATION. THE SILICONE EMPLOYED IN THIS PRODUCT IS A MEDICAL GRADE SILICONE AUTHORIZED FOR PRODUCT USE. SILICONE CONTENT TESTS ARE PERFORMED DURING THE MANUFACTURING PROCESS OF EACH LOT NUMBER ACCORDING TO PROCEDURE, RESULTS WERE REVIEWED FOR THE REPORTED LOT AND FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO DETERMINE A ROOT CAUSE AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT A "GREASY" LAYER OF FOREIGN MATTER WAS VISIBLE INSIDE THE BD PLASTIPAK¿ LUER-LOK¿ SYRINGE BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM DUTCH TO ENGLISH: WE HAVE BEEN USING THIS SYRINGE FOR A LONG TIME WITHOUT PROBLEMS, WITH ONE OF THE CYTOSTATIC FILLERS, MITOMYCIN, A GREASY LAYER (SILICONE?) IS VISIBLE ON THE INSIDE OF THE SYRINGE. THIS IS REALLY NOT ON THE OUTSIDE, NOT TO WIPE OR SCRAPE OFF. I CANNOT SEND THIS SYRINGE BECAUSE OF THE RISKY CONTENT. WE TOOK THE REST OF THIS BATCH OF SYRINGES (12 PIECES) FROM THE PHARMACY SHELVES AND THEN USED ANOTHER LOT NUMBER WITHOUT ANY PROBLEMS.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A "GREASY" LAYER OF FOREIGN MATTER WAS VISIBLE INSIDE THE BD PLASTIPAK¿ LUER-LOK¿ SYRINGE BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: WE HAVE BEEN USING THIS SYRINGE FOR A LONG TIME WITHOUT PROBLEMS, WITH ONE OF THE CYTOSTATIC FILLERS, MITOMYCIN, A GREASY LAYER (SILICONE?) IS VISIBLE ON THE INSIDE OF THE SYRINGE. THIS IS REALLY NOT ON THE OUTSIDE, NOT TO WIPE OR SCRAPE OFF. I CANNOT SEND THIS SYRINGE BECAUSE OF THE RISKY CONTENT. WE TOOK THE REST OF THIS BATCH OF SYRINGES (12 PIECES) FROM THE PHARMACY SHELVES AND THEN USED ANOTHER LOT NUMBER WITHOUT ANY PROBLEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
571314 BD PLASTIPAK LUER-LOK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 1912280

Patients

Seq Age Sex Outcome Treatment
1 Other