BD INSULIN SYRINGE WITH BD ULTRA-FINE NEEDLE
Report
- Report Number
- 1920898-2020-00610
- Event Type
- Malfunction
- Date Received
- June 2, 2020
- Date of Event
- May 22, 2020
- Report Date
- May 26, 2020
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 00382908438032
- PMA / PMN Number
- K024112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MEDICAL DEVICE EXPIRATION DATE: NA. INVESTIGATION SUMMARY: LEVEL A INVESTIGATION - COMPLAINT EVALUATION / COMPLAINT HISTORY CHECK FOR THE EVENT(S) THAT OCCURRED. SEVERITY: S_1__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR LABEL INFORMATION MISSING (EXPIRATION DATE) ON LOT # 7240810. INVESTIGATION SUMMARY: NO SAMPLES WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH # 7240810 ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200719497] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. THIS BATCH WAS MANUFACTURED BEFORE EXPIRATION DATE WAS PRINTED ON PACKAGING. INVESTIGATION CONCLUSION: AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT BD INSULIN SYRINGE WITH BD ULTRA-FINE¿ NEEDLE WAS MISSING LABEL INFORMATION. THIS WAS DISCOVERED BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 328438 BATCH NO. 7240810. IT WAS REPORTED THAT EXPIRATION DATE IS MISSING FROM BOX.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 571794 | BD INSULIN SYRINGE WITH BD ULTRA-FINE NEEDLE | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 328438 | 7240810 | 00382908438032 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |