FDA Adverse Event Malfunction Summary report: N

BENDING/CUTTING PLIERS

MDR report key: 10106122 · Received June 1, 2020

Report

Report Number
2939274-2020-02623
Event Type
Malfunction
Date Received
June 1, 2020
Date of Event
May 19, 2020
Report Date
May 19, 2020
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HTC
UDI-DI
10886982198525
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6- THE BENDING/CUTTING PLIERS (P/N:391.962, LOT #: T991566) WAS RETURNED AND RECEIVED AT US CQ. UPON VISUAL INSPECTION, THE CARBIDE INSERT WAS MISSING ON THE CUTTING SLOT OF THE DEVICE AND THE DEVICE WAS OBSERVED TO BE FUNCTIONING LOOSE. THERE WERE SCRATCHES OBSERVED ON THE DEVICE BUT HAS NO IMPACT ON THE FUNCTIONALITY OF THE DEVICE. NO OTHER ISSUES WERE OBSERVED WITH THE RETURNED COMPONENTS OF THE DEVICE. THE COMPLAINT CONDITION IS CONFIRMED FOR THE BENDING/CUTTING PLIERS (P/N:391.962, LOT #: T991566). THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE ISSUE AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE COULD BE DUE TO UNINTENDED FORCES APPLIED TO THE DEVICE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT PART NUMBER: 391.962, LOT NUMBER: T991566, MANUFACTURING SITE: TUTTLINGEN, RELEASE TO WAREHOUSE DATE: 07-JUN-2013. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: THE DEVICE HAS BEEN RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. REPORTER IS COMPANY REPRESENTATIVE. WITHOUT A LOT NUMBER, THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED AS NO PRODUCT WAS RECEIVED. THE INVESTIGATION COULD NOT BE COMPLETED, NO PRODUCT WAS RECEIVED; NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2020, A PART OF THE BENDING/CUTTING PLIERS CHIPPED OFF WHILE CUTTING THE 2.7 LOCKING COMPRESSION PLATE (LCP) ADAPTATION PLATE. THE CUTTING PLIERS SUCCESSFULLY CUT THE PLATE BUT CHIPPED IN THE PROCESS. THE PIECE WAS RECOVERED ON THE BACK TABLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT A SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. CONCOMITANT DEVICE REPORTED: UNKNOWN 2.7 LCP ADAPTATION PLATE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY 1) THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
570480 BENDING/CUTTING PLIERS PLIERS,SURGICAL HTC WRIGHTS LANE SYNTHES USA PRODUCTS LLC 391.962 T991566 10886982198525

Patients

Seq Age Sex Outcome Treatment
1 UNK - PLATES| UNK - PLATES