FDA Adverse Event
Injury
Summary report: N
RIATA PASSIVE FIXATION
MDR report key: 1010586
·
Received March 10, 2008
Report
- Report Number
- 2017865-2008-01227
- Event Type
- Injury
- Date Received
- March 10, 2008
- Date of Event
- December 31, 2007
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS FOUND AN INTERNAL INSULATION ABRASION UNDER THE SVC SHOCK COIL. ONE OF THE RV COILS WAS FRACTURED. THIS DAMAGE MOST LIKELY CAUSED THE RV CABLE AND SVC SHOCK COIL TO SHORT CIRCUIT WHICH CAUSED THE RV CABLE TO BE FRACTURED WHEN IT WAS EXPOSED TO SHOCK. EXTERNAL INSULATION ABRASION WAS FOUND AT 35.5 TO 36.8CM FROM THE CONNECTOR PIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RIATA PASSIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 1571/65 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |