FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 1010099 · Received March 7, 2008

Report

Report Number
1220908-2008-00447
Event Type
Malfunction
Date Received
March 7, 2008
Report Date
February 19, 2008
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO AN UNSEATED FLEX CABLE. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE DISPLAYED A "DEFIB FAULT 72" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION MSERIES NA

Patients

Seq Age Sex Outcome Treatment
1 NA