AXSYM DIGOXIN III
Report
- Report Number
- 2623532-2007-00163
- Event Type
- Malfunction
- Date Received
- May 11, 2007
- Date of Event
- January 12, 2007
- Report Date
- April 11, 2007
- Manufacturer
- ABBOTT DIAGNOSTICS INTERNATIONAL, LTD.
- Product Code
- KXT
- PMA / PMN Number
- K061249
- Removal / Correction Number
- 2623532-4/10/07-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- UNKNOWN
Narratives
THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE CUSTOMER STATES, THAT WHILE PERFORMING A CORRELATION STUDY BETWEEN THE AXSYM DIGOXIN II ASSAY VS. THE AXSYM DIGOXIN III ASSAY, THEY NOTED THAT PTS SAMPLES RESULTS GENERATED ON THE AXSYM DIGOXIN III WERE HIGHER COMPARED TO AXSYM DIGOXIN II. FOR EXAMPLE, A PT SAMPLE GENERATED A RESULT OF 1.08 NG/ML USING THE AXSYM DIGOXIN III WHICH IS HIGHER THAN THE RESULT OF 0.91 NG/ML USING THE AXSYM DIGOXIN II. THE CUSTOMER ALSO STATES, THAT THEY ARE SOMETIMES GETTING ERROR CODE 1118 (INVALID TEST RESULT, INTERCEPT TOO LOW) AND ERROR CODE 1062 (INVALID TEST RESULT, READ PRECISION) WHEN RUNNING A PT SAMPLE ON THE AXSYM DIGOXIN III. THERE WAS NO IMPACT TO PTS MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AXSYM DIGOXIN III | MEIA FOR THE MEASUREMENT OF DIGOXIN | KXT | ABBOTT DIAGNOSTICS INTERNATIONAL, LTD. | NA | 42722Q100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | AXSYM ANALYZER |