AFINION 2
Report
- Report Number
- 3003045237-2020-00005
- Event Type
- Malfunction
- Date Received
- May 29, 2020
- Date of Event
- April 16, 2020
- Report Date
- May 29, 2020
- Manufacturer
- ABBOTT DIAGNOSITCS TECHNOLOGIES AS
- Product Code
- JQT
- PMA / PMN Number
- K171650
- Removal / Correction Number
- 9613069-02/03/2020-001-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER
Narratives
COMPLAINTS HAVE BEEN RECEIVED REGARDING ELECTROSTATIC DISCHARGE (ESD) AND/OR POWER SUPPLY FAILURE FOR AFINION 2 AND ALERE AFINION AS100 ANALYZER. AN INVESTIGATION HAS CONCLUDED THAT THE ROOT CAUSE FOR THE OBSERVED LEVEL OF COMPLAINTS RELATED TO ESD EVENTS AND POWER SUPPLY FAILURE IS ASSOCIATED WITH THE POWER SUPPLY NOT WITHSTANDING BUILD-UP OF ELECTROSTATIC DISCHARGE IN THE USE ENVIRONMENT BASED ON THE DESIGN AND CONSTRUCTION OF THE POWER SUPPLY. AN 806 FOR ALERE AFINION AS100 ANALYZER AND AFINION 2 WAS SENT TO THE FDA ON 3 FEBRUARY 2020 FOR A CORRECTION OF THE POWER SUPPLY PROVIDED WITH THE ANALYZERS BY REPLACING THE CURRENT POWER SUPPLY WITH A NEW POWER SUPPLY CERTIFIED TO A HIGHER STANDARD. ADDITIONALLY AN UPDATE TO THE USER MANUALS FOR THE ANLAYZERS WAS MADE TO PROVIDE AWARENESS TO THE RISK OF ELECTROSTATIC DISCHARGE EVENTS. MANUFACTURER'S REFERENCE FOR THE 806 IS 9613069-02/03/2020-001-C (REFERENCE ASSIGNED BY FDA: (B)(4)).
THE CUSTOMER PLUGGED IN THE INSTRUMENT TO THE WALL SOCKET. THE CUSTOMER THEN NOTICED A SMALL FLASH AND HEARD A SOUND OF A SMALL ELECTRIC SHOCK AND SMELLED SMOKE. THE CUSTOMER NOTICED THE SMALL FLASH IN THE SOCKET AREA. SHE DESCRIBED IT AS SOMETHING AS STATIC SHOCK. SHE ALSO REPORTED THAT SHE "WAS PRETTY SURE" SHE SMELLED SOME MINOR SMOKE IN THE ROOM RIGHT AFTER THE INCIDENT THE CUSTOMER ALSO STATED THAT SHE TRIED TO START THE INSTRUMENT AFTER THE INCIDENT USING DIFFERENT CORDS AND POWER SOCKETS BUT THE INSTRUMENT DID NOT POWER ON AT ALL. THERE WAS NO INJURY OR HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 567453 | AFINION 2 | ANALYZER | JQT | ABBOTT DIAGNOSITCS TECHNOLOGIES AS | 1116553 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |