FDA Adverse Event Malfunction Summary report: N

SENSICA UO MONITOR ICU

MDR report key: 10095853 · Received May 28, 2020

Report

Report Number
1018233-2020-03474
Event Type
Malfunction
Date Received
May 28, 2020
Report Date
December 17, 2020
Manufacturer
ADAPTEC MEDICAL DEVICES LLC 3014271001
Product Code
EXS
UDI-DI
00854003008002
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED WAS CONFIRMED. THE ROOT CAUSE OF THE REPORTED ISSUE WAS USE-RELATED, AS THE USER IMPROPERLY REMOVED THE PATIENT RING CAUSING DAMAGE TO THE CONNECTION PINS. THE UNIT WAS EVALUATED UPON RECEIPT. LOAD CELL ASSEMBLY NEEDS TO BE REPLACED. PER EVALUATION, THE USER IMPROPERLY REMOVED THE PATIENT RING CAUSING DAMAGE TO THE CONNECTION PINS. NO REPAIRS WERE MADE, AS THE CUSTOMER DECLINED REPAIR AND REQUESTED THE UNIT BE SENT BACK WITHOUT SERVICE. AS A PART OF FIELD ACTION, THE FOLLOWING COMPONENTS WERE REPLACED/ADDED PER THE LATEST REVISION OF THE BILL OF MATERIAL FOR SENSICA ICU, PRODUCT CODE, REV H BASE PCBA, REV C SENSICA ICU BASE PCBA FIRMWARE, REV D SENSICA ICU APPLICATION SOFTWARE, REV A WINDOWS ENTERPRISE OS, REV A POWER SUPPLY LABEL. PRODUCT LABEL, AND USB CONNECTION LABEL WERE REPLACED WITH THE LATEST REVISION OF THE LABEL (REV C) PER THE LATEST BILL OF MATERIAL FOR SENSICA ICU, PRODUCT CODE SCCS1001. REPLACED PCBA, REV H WITH FIRMWARE - SERIAL NUMBER (B)(6). THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "TO ATTACH THE RING TO THE SENSICA UO SYSTEM STAND, TURN THE RING UPSIDE DOWN AND USE A CLOCKWISE MOTION TO TWIST AND "LOCK" THE RING ONTO THE RING INTERFACE. DO NOT APPLY EXCESSIVE FORCE OR TORQUE TO THE RING OR THE SYSTEM'S RING INTERFACE WHEN CONNECTING THE DEVICE TO AVOID DAMAGING COMPONENTS. DO NOT TURN THE RING COUNTERCLOCKWISE WHEN ATTACHING IT. " H11:SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SENSICA DEVICE HAD A LOAD CELL ISSUE. PER FOLLOW UP ON 28MAY2020, ENGINEER BRETT STATED THE SENSICA UNIT IS BEING RETURNED TO BD FOR REPAIRS.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SENSICA DEVICE HAD A LOAD CELL ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
558511 SENSICA UO MONITOR ICU SENSICA EXS ADAPTEC MEDICAL DEVICES LLC 3014271001 SCCS-1001 NA 00854003008002

Patients

Seq Age Sex Outcome Treatment
1