FDA Adverse Event Malfunction Summary report: N

SENSICA UO MONITOR ICU

MDR report key: 10095852 · Received May 28, 2020

Report

Report Number
1018233-2020-03472
Event Type
Malfunction
Date Received
May 28, 2020
Report Date
December 15, 2020
Manufacturer
ADAPTEC MEDICAL DEVICES LLC 3014271001
Product Code
EXS
UDI-DI
00854003008002
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED ISSUE WAS CONFIRMED. THE ROOT CAUSE OF THE REPORTED ISSUE WAS ISOLATED TO BE USER RELATED. USER IMPROPERLY REMOVED THE PATIENT RING CAUSING THE FRONT FACE OF THE LOAD CELL TO BREAK OFF COMPLETELY. LOAD CELL ASSEMBLY NEEDS TO BE REPLACED. CUSTOMER DECLINED ANY REPAIRS. AS A PART OF FIELD ACTION, COMPONENTS WERE REPLACED/ADDED: REV H BASE PCBA, REV C SENSICA ICU BASE PCBA FIRMWARE, REV D SENSICA ICU APPLICATION SOFTWARE, REV A WINDOWS ENTERPRISE OS, REV A POWER SUPPLY LABEL. PRODUCT LABEL, USB CONNECTION LABEL, WAS REPLACED WITH THE LATEST REVISION OF THE LABEL (REV C) PER THE LATEST BILL OF MATERIAL FOR SENSICA ICU, PRODUCT CODE SCCS1001. REPLACED PCBA REV H WITH FIRMWARE - SERIAL NUMBER (B)(6). PER THE LATEST REVISION OF THE BILL OF MATERIAL FOR SENSICA ICU, PRODUCT CODE SCCS1001. THE DEVICE HISTORY RECORD WAS REVIEWED AND FOUND NOTHING THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "TO ATTACH THE RING TO THE SENSICA UO SYSTEM STAND, TURN THE RING UPSIDE DOWN AND USE A CLOCKWISE MOTION TO TWIST AND "LOCK" THE RING ONTO THE RING INTERFACE. DO NOT APPLY EXCESSIVE FORCE OR TORQUE TO THE RING OR THE SYSTEM'S RING INTERFACE WHEN CONNECTING THE DEVICE TO AVOID DAMAGING COMPONENTS. DO NOT TURN THE RING COUNTERCLOCKWISE WHEN ATTACHING IT. "

Description of Event or Problem · 0

IT WAS REPORTED THAT THE SENSICA DEVICE HAD A LOAD CELL ISSUE. PER FOLLOW UP ON 28MAY2020, ENGINEER BRETT STATED THE SENSICA UNIT IS BEING RETURNED TO BD FOR REPAIRS.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SENSICA DEVICE HAD A LOAD CELL ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
558509 SENSICA UO MONITOR ICU SENSICA EXS ADAPTEC MEDICAL DEVICES LLC 3014271001 SCCS-1001 NA 00854003008002

Patients

Seq Age Sex Outcome Treatment
1