ENDOPATH ENDOSCOPIC MULTIFEED STAPLER

MDR report key: 100940 · Received June 25, 1997

Report

Report Number
1527736-1997-01413
Event Type
Malfunction
Date Received
June 25, 1997
Date of Event
May 28, 1997
Report Date
June 24, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GAG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY # 973325. VISUAL INSPECTIONS & RESULTS: HEAD ROTATES, YES; NOSE SHROUD CRACKED/BROKEN, YES; STAPLES IN NOSE, YES (2); STAPLES IN THE TRACK, YES (5) AND TRIGGER ENGAGED WITH PRECOCK, YES. FUNCTIONAL TESTS & RESULTS: CYCLED INSTRUMENT, NO. ANALYSIS CONCLUSION: BASED ON THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE INSTRUMENT JAMMED DURING SURGERY DUE TO A YIELDED CARTRIDGE NOSE WELD. NO CONCLUSION COULD BE REACHED AS TO HOW THE NOSE WELD HAD YIELDED. CO REVIEWS EACH INCIDENT AS IT OCCURS IN AN EFFORT TO CONTINUOUSLY IMPROVE OUR PRODUCTS AND PROCESSES.

Description of Event or Problem · 1

IT WAS REPORTED THE DEVICE WAS USED DURING AN UNKNOWN PROCEDURE. IT WAS REPORTED BY THE AFFILIATE THAT THE DEVICE JAMMED. THERE WAS NO PATIENT CONSEQUENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ENDOSCOPIC MULTIFEED STAPLER ENDOSCOPIC STAPLER GAG ETHICON ENDO-SURGERY, INC. NA K46659

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other