FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 1008997 · Received March 6, 2008

Report

Report Number
3002158293-2008-00088
Event Type
Malfunction
Date Received
March 6, 2008
Date of Event
February 15, 2008
Report Date
March 5, 2008
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT HAS BEEN COMPLETED. THE PROBLEM WAS CONFIRMED. UPON FURTHER INSPECTION, ELECTRODE BELT HAD PINS THAT WERE BENT INSIDE THE CONNECTOR. THE ROOT CAUSE OF THE BENT PINS WAS PROBABLY EXCESSIVE FORCE APPLIED TO THE CONNECTOR DURING MATING. THE CONNECTOR WAS FORCED INTO THE MONITOR WHICH DEFEATED THE CONNECTOR KEYING MECHANISM AND ALLOWED THE PINS TO MISALIGN WITH THE MATING SOCKETS. THE CONNECTOR NUT AND THE CONNECTOR BODY ALSO WERE DAMAGED. THE DAMAGED ELECTRODE BELT CONNECTOR WAS REPLACED. THE ELECTRODE BELT WAS RETESTED AND THEN RESTOCKED. NO ADVERSE EVENT RESULTED FROM THE BENT PINS. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A FEMALE PT CONTACTED LIFECOR CUSTOMER SUPPORT TO REPORT THAT SHE HAD DISCONNECTED THE ELECTRODE BELT FROM THE MONITOR AND UPON REINSERTING, IT SHE BROKE OFF SOME PINS. SUPPORT SENT A PT SERVICES REP (PSR) TO REPLACE THE ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3100 NA

Patients

Seq Age Sex Outcome Treatment
1 34 YR