NXSTAGE SYSTEM ONE
Report
- Report Number
- 3003464075-2020-00026
- Event Type
- Death
- Date Received
- May 22, 2020
- Date of Event
- April 3, 2020
- Report Date
- June 16, 2020
- Manufacturer
- NXSTAGE MEDICAL, INC.
- Product Code
- KDI
- PMA / PMN Number
- K050525
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NURSE
Narratives
THE DEVICE WAS RECEIVED AND EVALUATED AND WAS FOUND TO BE FUNCTIONING AS DESIGNED AND INTENDED. THE LOG FILE FOR THE TREATMENT WAS ANALYZED AND FINDINGS SHOWED THAT AT THE TIME OF THE EVENT THE PRIME AND ALARM TEST WAS STILL IN PROGRESS. DURING THE PRIME AND ALARMS TEST SOME SAFETY SYSTEMS, INCLUDING AIR ALARMS, ARE INACTIVE TO ALLOW THE USER TO EXPEL AIR FROM THE EXTRACORPOREAL CIRCUIT. THE NXSTAGE SYSTEM ONE USER GUIDE WARNS TO NEVER CONNECT THE PATIENT TO THE CARTRIDGE BEFORE YOU SEE THE TREATMENT PARAMETERS IN THE NXSTAGE SYSTEM ONE CYCLER¿S WINDOW, WHICH INDICATES THAT THE PRIME AND ALARMS TEST IS COMPLETED. THERE WAS NO DEVICE MALFUNCTION. THE AVAILABLE INFORMATION IS CONSISTENT WITH A USE ERROR AND THE DIAGNOSIS OF AIR EMBOLISM PROVIDED BY THE PHYSICIAN. H3 OTHER TEXT : PLACEHOLDER.
A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS CONDUCTED WHICH CONFIRMED THAT THE PRODUCT MET ALL QUALITY CRITERIA AND MANUFACTURING SPECIFICATIONS PRIOR TO RELEASE. (B)(4).
A REPORT WAS RECEIVED ON 24 APR 2020 FROM THE HOME THERAPY NURSE (HTN) REGARDING A (B)(6) YEAR OLD FEMALE WITH AN EXTENSIVE MEDICAL HISTORY WHO BECAME DIZZY AND UNRESPONSIVE AFTER TROUBLESHOOTING ALARMS AND TERMINATING HEMODIALYSIS TREATMENT WITH RINSEBACK ON (B)(6) 2020. 911 WAS CALLED AND THE PATIENT WAS TRANSPORTED TO HOSPITAL WHERE SHE SUBSEQUENTLY EXPIRED ON (B)(6) 2020. ADDITIONAL INFORMATION WAS RECEIVED 29 APR 2020 FROM THE HTN WHICH SHOWED RESUSCITATION MEASURES INCLUDED INTUBATION AND EPINEPHRINE (ROUTE NOT SPECIFIED). AFTER ARRIVING AT THE HOSPITAL THE PATIENT WAS ADMITTED TO THE INTENSIVE CARE UNIT (ICU). A CT SCAN REVEALED AN AIR EMBOLISM IN THE BRAIN. THE HOSPITAL COURSE INVOLVED THE PATIENT BEING MECHANICALLY VENTILATED FOR ACUTE RESPIRATORY FAILURE SECONDARY TO THE ANOXIC BRAIN INJURY AND AN ACUTE BIFRONTAL LOBE EMBOLIC CEREBROVASCULAR ACCIDENT (CVA). THE PATIENT REMAINED UNRESPONSIVE AND AREFLEXIC AND WAS GIVEN A POOR PROGNOSIS. THEIR CONDITION WORSENED DUE TO SYSTEMIC INFLAMMATORY RESPONSE SYNDROME, BRAIN INJURY, ACUTE ENCEPHALOPATHY AND SEIZURES. THE PATIENT WAS EXTUBATED ON (B)(6) 2020, PLACED ON COMFORT CARE AND EXPIRED AT 21:45HRS ON (B)(6) 2020 WITH CARDIAC ARREST REPORTED AS THE CAUSE OF DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 546920 | NXSTAGE SYSTEM ONE | HIGH PERMEABILITY HEMODIALYSIS SYSTEM | KDI | NXSTAGE MEDICAL, INC. | NX1000-1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Death| H |