FDA Adverse Event Malfunction Summary report: N

MONACO

MDR report key: 10082837 · Received May 22, 2020

Report

Report Number
1937649-2020-00011
Event Type
Malfunction
Date Received
May 22, 2020
Date of Event
April 30, 2020
Report Date
May 22, 2020
Manufacturer
ELEKTA INC.
Product Code
MUJ
PMA / PMN Number
K190178
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION WAS COMPLETED BY CONDUCTING A THOROUGH EVALUATION OF THE PRODUCT AND THE REPORTED INFORMATION. THE ISSUE WAS DETERMINED TO BE A DEFECT IN THE PRODUCT. BOLUS EXISTS IN THE REFERENCE PLAN AND IT IS DEFINED AS USER. DURING THE ATS ONLINE WORKFLOW, THE BOLUS IS CLEARED BY DESIGN TO FORCE THE USER TO RECREATE IT SINCE THE PATIENT OUTLINE MIGHT HAVE CHANGED. AFTER DOING THAT, THE PLAN IS CREATED, BUT THE BOLUS IS NOT ASSIGNED TO THE BEAMS IN THE BEAM SPREADSHEET FOR BOTH 3D AND IMRT. IN ADDITION, THE USER DOES NOT HAVE THE ABILITY TO REASSIGN THE BOLUS TO THE BEAM(S) AT THIS POINT. IF THE USER DOES NOT DETECT THAT BOLUS HAS NOT BEEN CONSIDERED IN THE DOSE CALCULATION, AND BOLUS IS ADDED DURING TREATMENT, THE DOSE DELIVERED WILL NOT MATCH THE DOSE DISPLAYED. UNDERDOSES COULD BE GREATER THAN 5%. THIS IS AN INTERNAL INVESTIGATION AT ELEKTA INC.AND THERE WERE NO PATIENTS INVOLVED. THE CORRECTIVE ACTIONS TAKEN AS A RESULT OF THIS INVESTIGATION IS THAT A PRODUCT BULLETIN WILL BE SENT AND THE ISSUE WILL BE RESOLVED IN THE MONACO 5.51.10 RELEASE.

Description of Event or Problem · 1

DURING AN INTERNAL INVESTIGATION AT ELEKTA INC., WHEN USING BOLUS ON THE REFERENCE PLAN, THE SYSTEM DETECTS THAT DURING THE ATS ONLINE WORKFLOW THE STRUCTURE BOLUS THAT WAS DEFINED AS 'USER' IS MISSING AND FORCES THE USER TO RECREATE THE BOLUS ONLINE. AFTER RECREATING THE BOLUS ONLINE, THE PLAN IS CREATED BUT THE BOLUS IS NOT ASSIGNED TO THE BEAMS IN THE BEAM SPREADSHEET FOR BOTH 3D AND IMRT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
546790 MONACO SYSTEM,PLANNING,RADIATION THERAPY TREATMENT MUJ ELEKTA INC.

Patients

Seq Age Sex Outcome Treatment
1