OUTLET PORT CLAMP
Report
- Report Number
- 1416980-2020-02863
- Event Type
- Malfunction
- Date Received
- May 21, 2020
- Date of Event
- April 26, 2020
- Report Date
- June 16, 2020
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FKK
- UDI-DI
- 05413760096186
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
ADDITIONAL INFORMATION IN D10, H3, H6. H10: TWO (2) ACTUAL SAMPLES WERE RECEIVED FOR EVALUATION. VISUAL INSPECTION DID NOT IDENTIFY ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. FUNCTIONAL TESTING WAS PERFORMED WITH NO ISSUES NOTED. THE REPORTED CONDITION WAS NOT VERIFIED. THE REMAINING TWO (2) DEVICES WERE NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
(B)(6). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT DURING USE OF 4 OUTLET PORT CLAMPS FLUID FLOW WAS OBSERVED WITH THE CLAMPS CLOSED. THIS OCCURRED DURING USE OF THE DEVICES FOR PERITONEAL DIALYSIS THERAPY. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 542247 | OUTLET PORT CLAMP | CLAMP, LINE | FKK | BAXTER HEALTHCARE CORPORATION | NA | 19C28H80 | 05413760096186 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |