OSS POROUS IM STEM 18.5 X 225
Report
- Report Number
- 0001825034-2020-02071
- Event Type
- Malfunction
- Date Received
- May 21, 2020
- Date of Event
- July 9, 2019
- Report Date
- May 21, 2020
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- KRO
- PMA / PMN Number
- K002757
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 0001825034-2020-00231. CONCOMITANT MEDICAL PRODUCTS: MOD ARTHRO 7 DEG LCK COLLAR CATALOG#: CP260602 LOT#: 968130, MOD ARTHRO NL 1CM DIASL CNCTR CATALOG#: CP260605 LOT#: 161950, OSS POROUS IM STEM 21.5 X 150 CATALOG#: 150400 LOT#: 360910, MOD ARTHRO NL 1CM DIASL CNCTR CATALOG#: CP260605 LOT#: 236860. REPORTED EVENT WAS CONFIRMED BY REVIEW OF MEDICAL RECORDS. MEDICAL RECORDS WERE PROVIDED AND REVIEWED BY A HEALTHCARE PROFESSIONAL. REVIEW FOUND THE STEM WAS WELL FIXED IN THE PATIENT. THE SURGEON TRIED TO REMOVE THE ALREADY FRACTURED DIAPHYSEAL CONNECTOR FROM THE STEM BUT COULDN¿T AND WAS NOTED AS AN INTRAOP COMPLICATION THAT DELAYED THE SURGERY. DEVICE HISTORY RECORD WAS REVIEWED AND NO DISCREPANCIES RELEVANT TO THE REPORTED EVENT WERE FOUND. ROOT CAUSE WAS UNABLE TO BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT DURING A REVISION OF A LEFT KNEE OSS SYSTEM DUE TO A BROKEN JUNCTION/CONNECTOR COMPONENT, THE SURGEON STATED THAT THE MORRIS TAPER COULD NOT BE BROKEN FROM THE FEMORAL STEM DESPITE MULTIPLE ATTEMPTS AND DIFFERENT INSTRUMENTS. THE SURGERY HAD TO BE MODIFIED TO CUT THE SEGMENTS APART, EXTENDING SURGICAL TIME BY AN HOUR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 544965 | OSS POROUS IM STEM 18.5 X 225 | PROSTHESIS, KNEE | KRO | ZIMMER BIOMET, INC. | N/A | 898500 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |