FDA Adverse Event Other Summary report: N

VASSCAN TABLE MODEL 2184

MDR report key: 1007971 · Received February 27, 2008

Report

Report Number
1932056-2008-00001
Event Type
Other
Date Received
February 27, 2008
Date of Event
December 11, 2006
Report Date
February 13, 2008
Manufacturer
MEDICAL POSITIONING, INC.
Product Code
LGX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INCIDENT RESULTED FROM WELD FAILURE. WHILE LESS THAN 0.02% WELDS HAVE FAILED, THE INCIDENCE OF FAILURE IN 2004 IS HIGHER THAN IN OTHER YEARS. FOR THAT REASON, ALL CUSTOMERS WITH THE SAME MODEL THAT WAS MANUFACTURED FROM THE SAME BATCH OF STEEL IN 2004 WILL BE CONTACTED AND AN UPGRADE WILL BE INSTALLED TO ENSURE NO FAILURES IN THE FUTURE.

Description of Event or Problem · 1

AS THE PT LAID ON THE VASCAN TABLE DURING AN EXAM, THE UPPER TRENDELENBURG ACTUATOR MOUNT BRACKET BROKE AWAY FROM THE FRAME. THIS CAUSED THE HEAD END OF THE TABLE TO FALL RAPIDLY TO THE FLOOR. THE PT FELL OFF THE TABLE AND LANDED ON THE FLOOR. PT COMPLAINED OF NECK PAIN. NO SUBSEQUENT INJURY HAS BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VASSCAN TABLE MODEL 2184 GENERAL VASCULAR ULTRASOUND TABLE LGX MEDICAL POSITIONING, INC. 2184

Patients

Seq Age Sex Outcome Treatment
1 Other