MAXPLUS POSITIVE PRESSURE CONNECTOR
Report
- Report Number
- 9616066-2020-01541
- Event Type
- Malfunction
- Date Received
- May 19, 2020
- Date of Event
- April 29, 2020
- Report Date
- April 29, 2020
- Manufacturer
- SISTEMAS MEDICOS ALARIS, S.A. DE C.
- Product Code
- FPA
- PMA / PMN Number
- K072542
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
A MP1000 CHINA PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION; HOWEVER THE CUSTOMER INDICATES THAT LEAKAGE WAS IDENTIFIED AT THE CONNECTION TO A PICC LINE. FURTHER INFORMATION PROVIDED BY THE CUSTOMER INDICATES THAT NO OBVIOUS DAMAGE WAS VISIBLE WHICH MAY HAVE CONTRIBUTED TO THE CUSTOMER'S EXPERIENCE. THE DETAILS OF THIS FEEDBACK WERE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. A REVIEW OF THE PRODUCTION RECORDS FOR LOT 16018067 DID NOT IDENTIFY ANY IN-PROCESS TESTING FAILURES OR QUALITY DEVIATIONS WHICH MAY HAVE RESULTED IN A REPORT OF THIS NATURE. THE ROOT CAUSE OF THE CUSTOMER¿S EXPERIENCE COULD NOT BE DETERMINED AS THE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION. IN THIS INSTANCE, WITHOUT A SAMPLE IT IS NOT POSSIBLE TO DETERMINE WHETHER A MANUFACTURING DEFECT COULD HAVE CAUSED OR CONTRIBUTED TO THE CUSTOMER¿S EXPERIENCE. A REVIEW OF THE CUSTOMER FEEDBACK DATABASE INDICATES THAT COMPLAINTS OF THIS NATURE ARE RARE AND THERE IS CURRENTLY NO TREND FOR ISSUES OF THIS NATURE AGAINST THE MP1000 CHINA PRODUCT.
IT WAS REPORTED THAT THE MAXPLUS POSITIVE PRESSURE CONNECTOR EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CONNECTOR WAS LEAKAGE, AND THE LIQUID WAS LEAKAGE FROM SCREW JOINT.
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: A MP1000 PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION; HOWEVER THE CUSTOMER INDICATES THAT A LEAKAGE WAS IDENTIFIED AT THE CONNECTION TO THE MALE LUER COMPONENT. FURTHER INFORMATION RELATING TO THE INFUSION SET UP CONFIRMED THE CONNECTING PRODUCTS WERE BARD PICC CATHETER AND A KDL LUER SLIP SYRINGE. ROOT CAUSE DESCRIPTION: DHR FOR THIS COMPLAINT: (B)(4), LOT 16018067, MODEL MP1000 CHINA, (B)(4), MFG DATE 1/17/2016. NO QN¿S OR DEVIATIONS DURING THE LOT MANUFACTURING. INVESTIGATION CONCLUSION: THE DETAILS OF THIS FEEDBACK WERE FORWARDED TO THE MANUFACTURING SITE FOR INVESTIGATION. A REVIEW OF THE PRODUCTION RECORDS FOR LOT 16018067 DID NOT IDENTIFY ANY IN-PROCESS TESTING FAILURES OR QUALITY DEVIATIONS WHICH MAY HAVE RESULTED IN A REPORT OF THIS NATURE. THE ROOT CAUSE OF THE CUSTOMER¿S EXPERIENCE COULD NOT BE DETERMINED AS THE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION. IN THIS INSTANCE, WITHOUT A SAMPLE IT IS NOT POSSIBLE TO DETERMINE WHETHER A MANUFACTURING DEFECT COULD HAVE CAUSED OR CONTRIBUTED TO THE CUSTOMER¿S EXPERIENCE. A REVIEW OF THE CUSTOMER FEEDBACK DATABASE INDICATES THAT THERE ARE A SMALL NUMBER OF SIMILAR REPORTS; HOWEVER, THESE COMPLAINTS HAVE NOT BEEN ATTRIBUTABLE TO A MAXPLUS PRODUCT DEFECT. MP1000 CHINA THIS IS AN INTERNATIONAL CODE- THE MODEL#/CATALOG# IDENTIFIED IS A CAREFUSION PRODUCT WHICH IS SAME OR SIMILAR TO A DEVICE THAT IS APPROVED FOR SALE DOMESTICALLY. THE DOMESTIC SIMILAR LIST NUMBER IS MP1000. THE 510K NUMBER PROVIDED IS FOR THE DOMESTIC SIMILAR PRODUCT.-K072542.
IT WAS REPORTED THAT THE MAXPLUS POSITIVE PRESSURE CONNECTOR EXPERIENCED LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE CONNECTOR WAS LEAKAGE, AND THE LIQUID WAS LEAKAGE FROM SCREW JOINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 537171 | MAXPLUS POSITIVE PRESSURE CONNECTOR | INTRAVASCULAR ADMINISTRATION SET | FPA | SISTEMAS MEDICOS ALARIS, S.A. DE C. | 16018067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |