NEXGEN COMPLETE KNEE SOLUTION LEGACY KNEE-POSTERIOR STABILIZED LPS-FLEX ARTICULA
Report
- Report Number
- 1822565-2008-00082
- Event Type
- Injury
- Date Received
- March 4, 2008
- Date of Event
- August 8, 2006
- Report Date
- February 6, 2008
- Manufacturer
- ZIMMER, INC.
- Product Code
- JWH
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
EVALUATION SUMMARY: LABORATORY EXAMINATION USING SEM WAS CONDUCTED ON THE SPINE FRAGMENT TO EVALUATE THE MATERIAL FAILURE PATTERN AND THE FRACTURE APPEARS TO HAVE OCCURRED BY FATIGUE. ALSO, THE CRACK APPEARS TO HAVE PROPAGATED FROM ANTERIOR TO POSTERIOR SIDE. THE ANTERIOR SURFACE ON THE SPINE FRAGMENT SHOWS MATERIAL DEFORMATION IN BOW-TIE SHAPE ON THE DISTAL END. FURTHER, THE MFG RECORDS SHOW THAT THE DEVICE WAS MANUFACTURED AND PACKAGED TO SPECS. IT'S LIKELY THAT THE SPINE FRACTURED DUE TO ABNORMAL AP ARTICULATION OF THE FEMORAL COMPONENT CAUSING ANTERIOR IMPACT AND MATERIAL FATIGUE BUT THE CAUSE COULD NOT BE DEFINITIVELY DETERMINED BASED ON THE AVAILABLE INFORMATION. ONLY THE FRACTURED SPINE WAS RETURNED FOR REVIEW. DEVICE HISTORY RECORDS INDICATE DEVICE MANUFACTURED TO SPECIFICATION. SCANNING ELECTRON MICROSCOPY ANALYSIS OF THE FRACTURE SURFACE WAS CONDUCTED ON THE BROKEN SPINE OF THE ARTICULATING SURFACE. THE ARTICULATING SURFACE WAS NOT RECEIVED FOR EXAMINATION. STRIATED FEATURES OBSERVED ON THE FRACTURE SURFACE INDICATED THAT FRACTURE OCCURRED BY FATIGUE. ENERGY DISPERSIVE SPECTROSCOPY ANALYSIS OF THE SPINE MATERIAL SHOWED A CARBON (C) PEAK IN THE SPECTRUM TYPICAL OF UHMWPE.
IT IS REPORTED BY THE PT THAT HE UNDERWENT TOTAL LEFT KNEE ARTHROPLASTY IN 2001. POST-OP, THE PT EXPERIENCED PAIN AND SYNOVITIS. THE PT WAS REVISED IN 2006, WHEREIN FRACTURE AND A SQUARE CENTIMETER LOOSE BODY OF THE TIBIAL COMPONENT WAS NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEXGEN COMPLETE KNEE SOLUTION LEGACY KNEE-POSTERIOR STABILIZED LPS-FLEX ARTICULA | KNEE PROSTHESIS | JWH | ZIMMER, INC. | NA | 69495300 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Hospitalization| R |