VASOVIEW HEMOPRO 2
Report
- Report Number
- 2242352-2020-00430
- Event Type
- Malfunction
- Date Received
- May 18, 2020
- Date of Event
- April 27, 2020
- Report Date
- May 18, 2020
- Manufacturer
- MAQUET CARDIOVASCULAR LLC
- Product Code
- GEI
- PMA / PMN Number
- K101274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
INTERNAL COMPLAINT NUMBER: (B)(4). THE DEVICE WAS RETURNED TO THE FACTORY ON (B)(6) 2020. AN INVESTIGATION WAS CONDUCTED ON (B)(6) 2020. A VISUAL INSPECTION WAS CONDUCTED. SIGNS OF CLINICAL USE AND EVIDENCE OF BLOOD WAS OBSERVED ON THE BTT. THE INSUFFLATION TUBE WAS OBSERVED TO BE INTACT, NO VISUAL DEFECTS WERE OBSERVED. THE BALLOON OF THE BTT WAS OBSERVED TO BE INTACT, NO VISUAL DEFECTS WERE OBSERVED. THE CLEAR PORT WAS OBSERVED TO BE DETACHED FROM THE BTT AND WAS RETURNED FOR EVALUATION. AN INFLATION TEST WAS UNABLE TO BE CONDUCTED DUE TO THE DETACHED PORT. BASED ON THE RETURNED CONDITION OF THE DEVICE, THE REPORTED FAILURE "BREAK" WAS CONFIRMED. THE CERTIFICATE OF CONFORMANCE (C OF C) WAS REVIEWED. THE VENDOR CERTIFIES THAT THIS DEVICE LOT CONFORMS TO ALL APPLICABLE PRODUCT SPECIFICATIONS.
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE USING VASOVIEW HEMOPRO 2, THE AIR INLET FOR THE BTT PORT BALLOON WAS BROKEN. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.
TRACKWISE ID # (B)(4). A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE REPORTED PRODUCT LOT NUMBER. THERE WERE NO NCMR¿S FOR THE REPORTED LOT NUMBER. THE DEVICE HAS NOT YET BEEN RETURNED TO MAQUET CARDIAC SURGERY FOR EVALUATION. WE ARE FOLLOWING UP WITH THE CUSTOMER FOR THE RETURN OF THE DEVICE. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED.
THE HOSPITAL REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE USING VASOVIEW HEMOPRO 2, THE AIR INLET FOR THE BTT PORT BALLOON WAS BROKEN. A REPLACEMENT DEVICE WAS USED TO COMPLETE THE PROCEDURE. THE HOSPITAL DID NOT REPORT ANY PATIENT EFFECTS.