FDA Adverse Event Malfunction Summary report: N

BD PRECISIONGLIDE¿ NEEDLE

MDR report key: 10063775 · Received May 15, 2020

Report

Report Number
1911916-2020-00466
Event Type
Malfunction
Date Received
May 15, 2020
Date of Event
April 20, 2020
Report Date
May 5, 2020
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMI
UDI-DI
30382903051107
PMA / PMN Number
K021475
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION SUMMARY: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. INVESTIGATION CONCLUSION: BASED ON NO SAMPLE, THE INVESTIGATION CONCLUDED, BD WAS NOT ABLE TO VERIFY THE INDICATED FAILURE. THIS IS THE 1ST COMPLAINT FOR LOT # 9212458 FOR THIS TYPE OF DEFECT OR SYMPTOM. THERE WAS NO DOCUMENTATION FOR THIS TYPE OF DEFECTS DURING THE ENTIRE PRODUCTION RUN OF THIS BATCH. ROOT CAUSE DESCRIPTION: NO ROOT CAUSE CAN BE DETERMINED AS NO SAMPLES WERE RECEIVED. RATIONALE: CAPA NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT "PLASTIC-LIKE THREADS" WERE SEEN IN THE BD LUER-LOK¿ DISPOSABLE SYRINGE WITH BD LUER-LOK¿ DURING USE AFTER IT WAS FILLED WITH INSULIN. THIS COMPLAINT WAS CREATED TO CAPTURE THE 1ST OF 2 RELATED INCIDENTS. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "CUSTOMER REPORTED PLASTIC-LIKE THREADS IN HER INSULIN AFTER HAVING FILLED IT INTO THE SYRINGE."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
524438 BD PRECISIONGLIDE¿ NEEDLE HYPODERMIC SINGLE LUMEN NEEDLE FMI BD MEDICAL (BD WEST) MEDICAL SURGICAL 305110 9212458 30382903051107

Patients

Seq Age Sex Outcome Treatment
1 26 YR Other