Tablo(R) Hemodialysis System

MDR report key: 10062576 · Received May 15, 2020

Report

Report Number
3010355846-2020-00019
Event Type
Death
Date Received
May 15, 2020
Date of Event
May 2, 2020
Report Date
May 15, 2020
Manufacturer
OUTSET MEDICAL, INC.
Product Code
KDI
UDI-DI
00850001011112
PMA / PMN Number
K160866
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

OUTSET MEDICAL, INC. FIELD SERVICE ENGINEER (FSE) HAS REVIEWED SITE SYSTEM LOGS WITH A PROCEDURE DATE OF (B)(6) 2020. NO RELATED SYSTEM ALARMS WERE FOUND TO HAVE OCCURRED DURING TREATMENT. THE DEVICE IS FUNCTIONING AS INTENDED POST TREATMENT. A REVIEW OF PRODUCTION RECORDS FOR THIS UNIT DID NOT NOTE ANY MANUFACTURING NONCONFORMANCES THAT WOULD CONTRIBUTE TO A PRODUCT EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT EXPIRED DURING TREATMENT. THE NURSES STATED THAT THE PATIENT WAS VERY UNSTABLE PRIOR TO STARTING THE TREATMENT, AND THEY HAD EXPECTED THAT HE WOULD NOT SURVIVE THE REST OF THE DAY BUT PERFORMED DIALYSIS AS A LAST EFFORT TO EXTEND HIS LIFE. THE DEATH IS BELIEVED TO BE RELATED TO THE PATIENT'S PRE-EXISTING CONDITIONS AND NOT DUE TO THE TABLO DEVICE. NO FURTHER INFORMATION WAS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
524305 Tablo(R) Hemodialysis System DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI OUTSET MEDICAL, INC. TABLO X HEMODIALYSIS SYSTEM 00850001011112

Patients

Seq Age Sex Outcome Treatment
1 Death