T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY
Report
- Report Number
- 3013756811-2020-47289
- Event Type
- Injury
- Date Received
- May 15, 2020
- Date of Event
- April 22, 2020
- Report Date
- May 15, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OZO
- UDI-DI
- 00850006613380
- PMA / PMN Number
- P180008
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SC, US
- Reporter Occupation
- OTHER
Narratives
TANDEM QUALITY ENGINEER EVALUATED PUMP DATA AND CONCLUDED THE FOLLOWING: A TOTAL OF 6 LOW EVENTS WERE DETECTED WITHIN THE SPECIFIED DATE AND TIME RANGE. A LOW EVENT IS DEFINED AS A GLUCOSE LESS THAN 54 MG/DL IN A 2-HOUR PERIOD. THE RESULTS OF THE INVESTIGATION FOR EACH EVENT ARE INCLUDED BELOW. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF "LOW" (INDICATING BELOW 40 MG/DL) TO 49 MG/DL WERE RECORDED AT 8:32:17 PM TO 9:02:17 PM ON 4/22/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 55 WAS ENUNCIATED AT 8:27:17 PM ON 4/22/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF "LOW" (INDICATING BELOW 40 MG/DL) TO 53 MG/DL WERE RECORDED AT 12:57:17 AM TO 02:57:16 AM ON 4/23/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR READING OF 52 WAS ENUNCIATED AT 12:57:17 AM ON 4/23/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF 49 TO 53 MG/DL WERE RECORDED AT 06:22:16 AM TO 06:57:16 AM ON 4/23/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 55 WAS ENUNCIATED AT 06:17:16 AM ON 4/23/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF 50 TO 53 MG/DL WERE RECORDED AT 5:02:17 PM TO 5:17:16 PM ON 4/23/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 52 WAS ENUNCIATED AT 5:02:17 PM ON 4/23/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF "LOW" (INDICATING BELOW 40 MG/DL) TO 52 MG/DL WERE RECORDED AT 03:52:15 AM TO 04:42:15 AM ON 4/24/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 52 WAS ENUNCIATED AT 03:52:15 AM ON 4/24/2020. THE USER MADE THE FOLLOWING BOLUS REQUEST(S) WITHOUT ENTERING BLOOD GLUCOSE: TIME: 10:36:43 PM DATE: 4/23/2020 IOB: 0.00 BG: 0 MG/DL REQUESTING A BOLUS WITH ENTERING BLOOD GLUCOSE MAY LEAD TO A LOW GLUCOSE EVENT. CONTINUOUS GLUCOSE MONITOR (CGM) VALUES OF "LOW" (INDICATING BELOW 40 MG/DL) TO 53 MG/DL WERE RECORDED AT 07:32:15 AM TO 10:57:15 AM ON 4/24/2020. A CGM ALERT 1 (CGM LOW ALERT) FOR A READING OF 55 WAS ENUNCIATED AT 07:27:15 AM ON 4/24/2020. THE CAUSE OF THE LOW BG COULD NOT BE DETERMINED BASED ON THE REVIEW CONDUCTED. THERE IS NO EVIDENCE THAT THE PUMP EXPERIENCED A MALFUNCTION OR FAILURE.
THE DEVICE IS EXPECTED TO BE RETURNED; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF THE DEVICE IS RECEIVED.
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED A LOW BLOOD GLUCOSE (BG) LEVEL OF BELOW 40 MG/DL, CAUSE WAS UNKNOWN. CUSTOMER CONSUMED LIQUID GLUCOSE AND ORANGE JUICE TO ADDRESS THE LOW BG. RECOMMENDATION WAS MADE TO CONSULT WITH HEALTHCARE PROVIDER REGARDING DIABETES MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 526157 | T:SLIM X2 INSULIN PUMP WITH BASAL-IQ TECHNOLOGY | AUTOMATED INSULIN DOSING, THRESHOLD SUSPEND | OZO | TANDEM DIABETES CARE | 1002717 | 00850006613380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Other | INSULIN: NOVOLOG| INSULIN: NOVOLOG |