FDA Adverse Event Injury Summary report: N

EQUINOXE

MDR report key: 10060718 · Received May 15, 2020

Report

Report Number
1038671-2020-00293
Event Type
Injury
Date Received
May 15, 2020
Date of Event
May 9, 2020
Report Date
June 4, 2020
Manufacturer
EXACTECH, INC.
Product Code
KWT
UDI-DI
10885862086693
PMA / PMN Number
K063569
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

THE REVISION REPORTED MAY HAVE BEEN THE RESULT OF NOT PROPERLY ALIGNING THE GLENOSPHERE RELATIVE TO THE GLENOID BASEPLATE DURING THE INITIAL SURGERY. THIS MISALIGNMENT LIKELY PROHIBITED THE GLENOSPHERE FROM FULLY SEATING AND THE GLENOSPHERE LOCKING SCREW FROM FULLY ENGAGING THE GLENOID BASEPLATE, RESULTING IN DISASSEMBLY OF THE GLENOSPHERE LOCKING SCREW. HOWEVER, THIS CANNOT BE CONFIRMED AS THE EXPLANTS WERE NOT AVAILABLE FOR EVALUATION. SECTION H11: THE FOLLOWING SECTIONS HAVE CORRECTED INFORMATION: (D4) CATALOG NUMBER: 320-15-05, SERIAL NUMBER: (B)(6), EXPIRATION DATE: 07-JUL-2024, UNIQUE IDENTIFIER (UDI) #:(B)(4). (D11) CONCOMITANT DEVICE(S): 320-42-00, 6090294 - EQUINOXE REVERSE 42MM HUMERAL LINER +0. (H4) DEVICE MANUFACTURE DATE: 10-JUL-2019 (H6) EVALUATION CODE: 3165, 2907.

Additional Manufacturer Narrative · 1

PENDING EVALUATION. CONCOMITANT MEDICAL DEVICE(S): 320-15-05, (B)(4), EQ REV LOCKING SCREW.

Description of Event or Problem · 1

AS REPORTED, APPROXIMATELY 9 MONTHS POSTOP THE INITIAL IMPLANT, THE (B)(6) Y/O PATIENT PRESENTED FOR THE POSTOPERATIVE CHECK OF THE LEFT SHOULDER. UPON EXAM, IT WAS NOTED THAT THE LOCKING SCREW HAD BACKED OUT SOMEHOW. A REVISION WAS SCHEDULED AND COMPLETED. THE LOCKING SCREW AND HUMERAL LINER WERE REPLACED. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. DEVICES DISPOSED OF BY HOSPITAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
527437 EQUINOXE REVERSE 42MM HUMERAL LINER +0 KWT EXACTECH, INC. 320-42-00 UNK 10885862086693

Patients

Seq Age Sex Outcome Treatment
1 75 YR Hospitalization| R