KYPHON HV-R BONE CEMENT
Report
- Report Number
- 2953769-2020-00007
- Event Type
- Injury
- Date Received
- May 15, 2020
- Date of Event
- April 15, 2020
- Report Date
- May 15, 2020
- Manufacturer
- MDT KYPHON NEUCHATEL MFG
- Product Code
- NDN
- PMA / PMN Number
- SEE H10
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
OTHER - CEMENT LEAKED TO SOMATIC ARTERIES. THIS PRODUCT IS NOT APPROVED FOR SALE IN US BUT A SIMILAR DEVICE WITH CATALOG # C01A, 510K# K180700 AND UDI (B)(4) IS APPROVED FOR SALE IN US. NEITHER THE DEVICE NOR FILMS OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE, WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE PATIENT UNDERWENT BALLOON KYPHOPLASTY SURGERY AT L2 AFTER 1 WEEK OF COMPRESSION FRACTURE AND PRIMARY OSTEOPOROSIS. INTRA-OP, AFTER CEMENT FILLING WAS COMPLETED, CHECKING WAS PERFORMED WITH C-ARM. AFTER THAT, CEMENT SHADOW OF ABOUT 1CM WAS CHECKED ON THE LEFT SIDE OF THE VERTEBRAL BODY. WHEN RADIOGRAPH WAS CHECKED AFTER THE OPERATION, CEMENT SHADOW WAS CHECKED ON THE BLOOD VESSEL WHICH WAS CONSIDERED TO BE SOMATIC ARTERIES. ABOUT 10 MINUTES AFTER CEMENT STIRRING, CEMENT WAS TAPPED WITH FINGERS AND THREADS COULD NOT BE SEEN ANYMORE, AND IT WAS JUDGED THAT THE CEMENT VISCOSITY WAS APPROPRIATE. MOREOVER, IT WAS CHECKED THAT THE CEMENT DID NOT LEAK TO THE VERTEBRAL BODY ENDPLATE, AND DID NOT LEAK TO THE ANTERIOR AND POSTERIOR SIDE AS WELL. ON THE FOLLOWING DAY OF THE OPERATION DATE, THE PATIENT¿S SYMPTOMS DUE TO COMPRESSION FRACTURE IMPROVED AND THE PATIENT WAS DISCHARGED FROM THE HOSPITAL. THERE WAS NO PATIENT'S CONDITION WORSEN AFTER THAT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 526456 | KYPHON HV-R BONE CEMENT | CEMENT, BONE, VERTEBROPLASTY | NDN | MDT KYPHON NEUCHATEL MFG | NA | EL70129 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |