FDA Adverse Event Malfunction Summary report: N

R SERIES DEFIBRILLATOR

MDR report key: 10057855 · Received May 14, 2020

Report

Report Number
1220908-2020-01299
Event Type
Malfunction
Date Received
May 14, 2020
Report Date
April 23, 2020
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
UDI-DI
00847946017309
PMA / PMN Number
K060559/P160
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO A LOOSE MFE SHIELD ON THE ANALOG BOARD. THE RETAINER SNAP HOOKS FAILED TO STICK TO THE BOARD HOLDING THE MFE SHIELD ON. THE ANALOG BOARD WAS REPLACED TO RESOLVE THE REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "ECG DISABLED" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
521368 R SERIES DEFIBRILLATOR DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION R SERIES NA 00847946017309

Patients

Seq Age Sex Outcome Treatment
1