FDA Adverse Event Malfunction Summary report: N

T:SLIM INSULIN DELIVERY SYSTEM

MDR report key: 10053986 · Received May 13, 2020

Report

Report Number
3013756811-2020-46065
Event Type
Malfunction
Date Received
May 13, 2020
Date of Event
April 17, 2020
Report Date
May 13, 2020
Manufacturer
TANDEM DIABETES CARE
Product Code
LZG
UDI-DI
00852162004545
PMA / PMN Number
K111210
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PER TANDEM¿S CARTRIDGE INSTRUCTIONS FOR USE: "REPLACE THE CARTRIDGE EVERY 48 HOURS IF USING HUMALOG; EVERY 72 HOURS IF USING NOVOLOG." NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT MULTIPLE INTERMITTENT OCCLUSION ALARMS OCCURRED. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 160-287 MG/DL. REPORTEDLY, THE PUMP SUPPLIES WERE CHANGED TO ADDRESS THE ISSUE. ADDITIONALLY, THE PUMP SUPPLIES WERE IN USE FOR MORE THAN 3 DAYS WITH HUMALOG INSULIN. TANDEM TECHNICAL SUPPORT EDUCATED CUSTOMER REGARDING CARTRIDGE/INSULIN LABELING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
517246 T:SLIM INSULIN DELIVERY SYSTEM INSULIN PUMP LZG TANDEM DIABETES CARE 004628 00852162004545

Patients

Seq Age Sex Outcome Treatment
1 78 YR