FDA Adverse Event Injury Summary report: N

TANDEMLUNG OXYGENATOR

MDR report key: 10051759 · Received May 13, 2020

Report

Report Number
2531527-2020-00016
Event Type
Injury
Date Received
May 13, 2020
Date of Event
August 10, 2018
Report Date
April 13, 2020
Manufacturer
CARDIACASSIST INC.
Product Code
DTZ
PMA / PMN Number
K153295
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.10: THROUGH FOLLOW-UP COMMUNICATION LIVANOVA LEARNED THAT THE ISSUE WAS RESOLVED DURING THE PROCEDURE BY STOPPING THE PUMP FOR ONE MINUTE AND EXCHANGING THE OXYGENATOR. THE SUPPORT WAS THUS STARTED AGAIN. CLOTTING IS A MULTIFACTORIAL EVENT THAT MAY BE RELATED TO INSUFFICIENT ANTICOAGULATION, EXTENDED PERIODS OF LOW FLOW AND PATIENT COAGULOPATHIES. BASED ON THE AVAILABLE INFORMATION, THE EXACT CAUSE OF THE THROMBUS FORMATION IS THIS CASE CANNOT BE DETERMINED. IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 0

SEE INITIAL REPORT.

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. LOT NUMBER IS UNKNOWN. IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT IF AVAILABLE. CARDIAC ASSIST INC. MANUFACTURES THE TUNDEMLUNG OXYGENATOR. THE INCIDENT OCCURRED IN (B)(6). IF ANY ADDITIONAL INFORMATION PERTINENT TO THE REPORTED EVENT IS RECEIVED, IT WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

LIVANOVA RECEIVED A REPORT THAT DURING A SUPPORT, AFTER TANDEMHEART INSERTION, A THROMBUS FORMATION WAS OBSERVED IN THE CIRCUIT AT THE PROXIMAL SIDE OF THE TUNDEMLUNG OXYGENATOR. THE TANDEMHEART COMPONENT AND THE OXYGENATOR WERE EXCHANGED. THERE WAS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
514658 TANDEMLUNG OXYGENATOR OXYGENATOR, CARDIOPULMONARY BYPASS DTZ CARDIACASSIST INC. 5160-0000

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention