FDA Adverse Event Injury Summary report: N

POWER X1

MDR report key: 10048655 · Received May 12, 2020

Report

Report Number
1523574-2020-00014
Event Type
Injury
Date Received
May 12, 2020
Date of Event
May 3, 2020
Report Date
July 24, 2020
Manufacturer
FERNO-WASHINGTON, INC.
Product Code
FPO
UDI-DI
00190790002183
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE STRETCHER WAS RETURNED TO MANUFACTURER FOR EVALUATION. A VISUAL AND FUNCTIONAL EVALUATION WAS CONDUCTED AND IT WAS DISCOVERED AN IMPROPERLY INSTALLED STOW NET WAS INTERFERING WITH THE LOAD FRAME RELEASE BAR. THIS INTERFERENCE MAY HAVE NOT ALLOWED THE LOAD FRAME TO LOCK INTO POSITION WHICH WOULD CREATE IMPROPER WEIGHT DISTRIBUTION ON THE LOAD BEARING COMPONENTS AND COULD RESULT IN THE REPORTED ISSUE. THE STRETCHER WAS REPAIRED AND RETURNED TO THE CUSTOMER. THE IFU FOR THE STOW NET INCLUDES CLEAR INSTRUCTION ON PROPER INSTALLATION ON THE STRETCHER. NO ADDITIONAL INFORMATION WAS PROVIDED ON THE ALLEGED MEDIC INJURY.

Description of Event or Problem · 0

THE COMPLAINT ALLEGES WHILE UNLOADING A 430LB PATIENT FROM THE AMBULANCE, THE DROP FRAME ALLEGEDLY FAILED CAUSING THE WEIGHT OF THE STRETCHER TO SHIFT TOWARD THE OPERATORS. THE STRETCHER DID NOT FALL AND THE PATIENT WAS NOT INJURED; HOWEVER, A MEDIC IS ALLEGING A BACK AND KNEE INJURY, AS A RESULT. THE MEDIC WAS EVALUATED BY THE COMPANY'S CORPORATE HEALTH REPRESENTATIVE. NO FURTHER INFORMATION HAS BEEN PROVIDED.

Description of Event or Problem · 1

THE COMPLAINT ALLEGES WHILE UNLOADING A (B)(6) PATIENT FROM THE AMBULANCE, THE DROP FRAME ALLEGEDLY FAILED CAUSING THE WEIGHT OF THE STRETCHER TO SHIFT TOWARD THE OPERATORS. THE STRETCHER DID NOT FALL AND THE PATIENT WAS NOT INJURED; HOWEVER, A MEDIC IS ALLEGING A BACK AND KNEE INJURY, AS A RESULT. THE MEDIC WAS EVALUATED BY THE COMPANY'S CORPORATE HEALTH REPRESENTATIVE. NO FURTHER INFORMATION HAS BEEN PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
513736 POWER X1 POWER XI FPO FERNO-WASHINGTON, INC. 015807 00190790002183

Patients

Seq Age Sex Outcome Treatment
1 Other