T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY
Report
- Report Number
- 3013756811-2020-46105
- Event Type
- Malfunction
- Date Received
- May 8, 2020
- Date of Event
- April 17, 2020
- Report Date
- May 8, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00850006613205
- PMA / PMN Number
- DEN180058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
THE T:SLIM X2 WITH CIQ TECHNOLOGY USER GUIDE STATES: DO NOT USE ANY OTHER INSULIN WITH YOUR SYSTEM OTHER THAN U-100 HUMALOG OR U-100 NOVOLOG. ONLY HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE VERY HIGH OR A VERY LOW BLOOD GLUCOSE. TECHNICAL SUPPORT INFORMED CUSTOMER THAT IS OFF-LABEL PER THE USER GUIDE. PER TANDEM¿S CARTRIDGE INSTRUCTIONS FOR USE: ¿MAKE SURE THAT INSULIN IS AT ROOM TEMPERATURE BEFORE USE OR AIR BUBBLES COULD FORM IN THE CARTRIDGE.¿ NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT AN OCCLUSION ALARM OCCURRED. CUSTOMER REPORTED USING APIDRA INSULIN. ADDITIONALLY, CUSTOMER REPORTED USING COLD INSULIN. TANDEM CUSTOMER TECHNICAL SUPPORT INFORMED CUSTOMER THAT APIDRA IS OFF LABEL PER THE USER GUIDE. PER TANDEM¿S CARTRIDGE INSTRUCTIONS FOR USE: ¿MAKE SURE THAT INSULIN IS AT ROOM TEMPERATURE BEFORE USE OR AIR BUBBLES COULD FORM IN THE CARTRIDGE.¿ CUSTOMER CHANGED TUBING TO ADDRESS THE OCCLUSION ALARMS AND SUCCESSFULLY RESUMED INSULIN. CUSTOMER'S BLOOD GLUCOSE WAS 175 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 504108 | T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1002717 | 00850006613205 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | INSULIN: APIDRA, INFUSION SET: AUTOSOFT XC |