FDA Adverse Event Other Summary report: N

BD CRITIKIT IN-LINE ICE TEMPERATURE INJECTATE KIT

MDR report key: 1003826 · Received February 26, 2008

Report

Report Number
8020616-2008-00011
Event Type
Other
Date Received
February 26, 2008
Date of Event
January 29, 2008
Report Date
February 17, 2008
Manufacturer
BECTON DICKINSON CRITICAL CARE SYSTEMS PTE LTD.
Product Code
DRS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER. CONCLUSIONS: A ROOT CAUSE ANALYSIS COULD NOT BE DETERMINED AS THE SAMPLE WAS NOT SENT FOR EVALUATION. A REVIEW OF THE MANUFACTURING PROCESS SHOWED THAT THE TRANSDUCER ASSEMBLY GOES THROUGH 100% VISUAL INSPECTION TO CHECK FOR CRACKED OR DAMAGED COMPONENTS DURING SUBASSEMBLY PROCESS. THE KIT ASSEMBLY ALSO GOES THROUGH 100% VISUAL INSPECTION TO CHECK FOR DAMAGED ASSEMBLY DURING THE KIT MANUFACTURING PROCESS AND SAMPLING VISUAL INSPECTION FOR DAMAGED COMPONENT DURING OUTGOING INSPECTION. THE ENGINEER CONCLUDES THAT A POSSIBLE CAUSE OF THIS EVENT COULD BE THE USER ACCIDENTALLY BENT THE SYRINGE DURING APPLICATION THAT CAUSED THE SYRINGE TO SNAP OFF. (B) (4). (B) (4).

Description of Event or Problem · 1

THE SYRINGE BROKE OFF FROM THE TUBING WHEN THE NURSE TRIED TO USE IT FOR THE SECOND INJECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD CRITIKIT IN-LINE ICE TEMPERATURE INJECTATE KIT PRESSURE MONITORING ACCESSORY DRS BECTON DICKINSON CRITICAL CARE SYSTEMS PTE LTD. NA UNK

Patients

Seq Age Sex Outcome Treatment
1 UNK