BD CRITIKIT IN-LINE ICE TEMPERATURE INJECTATE KIT
Report
- Report Number
- 8020616-2008-00011
- Event Type
- Other
- Date Received
- February 26, 2008
- Date of Event
- January 29, 2008
- Report Date
- February 17, 2008
- Manufacturer
- BECTON DICKINSON CRITICAL CARE SYSTEMS PTE LTD.
- Product Code
- DRS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- UNKNOWN
Narratives
ADDITIONAL INFORMATION HAS BEEN REQUESTED FROM THE CUSTOMER. CONCLUSIONS: A ROOT CAUSE ANALYSIS COULD NOT BE DETERMINED AS THE SAMPLE WAS NOT SENT FOR EVALUATION. A REVIEW OF THE MANUFACTURING PROCESS SHOWED THAT THE TRANSDUCER ASSEMBLY GOES THROUGH 100% VISUAL INSPECTION TO CHECK FOR CRACKED OR DAMAGED COMPONENTS DURING SUBASSEMBLY PROCESS. THE KIT ASSEMBLY ALSO GOES THROUGH 100% VISUAL INSPECTION TO CHECK FOR DAMAGED ASSEMBLY DURING THE KIT MANUFACTURING PROCESS AND SAMPLING VISUAL INSPECTION FOR DAMAGED COMPONENT DURING OUTGOING INSPECTION. THE ENGINEER CONCLUDES THAT A POSSIBLE CAUSE OF THIS EVENT COULD BE THE USER ACCIDENTALLY BENT THE SYRINGE DURING APPLICATION THAT CAUSED THE SYRINGE TO SNAP OFF. (B) (4). (B) (4).
THE SYRINGE BROKE OFF FROM THE TUBING WHEN THE NURSE TRIED TO USE IT FOR THE SECOND INJECTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BD CRITIKIT IN-LINE ICE TEMPERATURE INJECTATE KIT | PRESSURE MONITORING ACCESSORY | DRS | BECTON DICKINSON CRITICAL CARE SYSTEMS PTE LTD. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |