FDA Adverse Event Injury Summary report: N

OXYGEN ANALYZER

MDR report key: 1003785 · Received February 25, 2008

Report

Report Number
1722070-2008-00001
Event Type
Injury
Date Received
February 25, 2008
Date of Event
January 22, 2008
Report Date
February 25, 2008
Manufacturer
MAXTEC, INC.
Product Code
CCL
PMA / PMN Number
K911344
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION FOR THIS REPORT IS FROM USER FACILITY. EVENT WAS UNKNOWN TO MAXTEC, UNTIL USER FACILITY FAXED A COPY OF THE MDR TO US ON 02/01/2008. THE PRODUCT HAS NOT BEEN ANALYZED YET. SALES REPRESENTATIVE FROM MAXTEC, WAS ALLOWED TO VIEW PRODUCT AND TAKE PICTURES; IN APPROXIMATELY TWO WEEKS MAXTEC ENGINEERS WILL BE ALLOWED TO VIEW PRODUCT IN USER FACILITY, AND ANALYZE PRODUCT.

Description of Event or Problem · 1

INFANT PATIENT WAS RESTING IN A BASSINET UNDER AN INFANT WARMER WITH AN OXYGEN HOOD IN PLACE. OXYGEN HOOD WAS PLACED OVER THE INFANT'S HEAD AND WAS ATTACHED TO AN OXYGEN AIR BLENDER FOR ADMINISTRATION OF 100% OXYGEN. AS THE STAFF APPROACHED THE BASSINET, THEY OBSERVED A FLASH OF FLAMES UNDER THE OXYGEN HOOD. STAFF REMOVED HOOD AND REMOVED OXYGEN FROM THE WALL. STAFF USED A BLANKET TO EXTINGUISH THE FIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OXYGEN ANALYZER OM25AE CCL MAXTEC, INC. OM25AE GC0228991

Patients

Seq Age Sex Outcome Treatment
1 12 HR Hospitalization| L| R