LIFEPAK 20 DEFIBRILLATOR/MONITOR

MDR report key: 1003747 · Received February 22, 2008

Report

Report Number
3015876-2008-00086
Event Type
Malfunction
Date Received
February 22, 2008
Date of Event
January 25, 2008
Report Date
January 25, 2008
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
LDD
PMA / PMN Number
K012274
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT THE DISPLAY WAS BLANK AND THE SYSTEM WOULD NOT TURN ON. PHYSIO REPLACED THE USER INTERFACE PCB ASSEMBLY AND THEN OBSERVED THAT IT WOULD PROPERLY POWER UP. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.

Description of Event or Problem · 1

FOUND DURING TESTING. ACCORDING TO THE REPORTER, THE DISPLAY WAS BLANK AND THE DEVICE WOULD NOT TURN ON. THERE WAS NO PATIENT INVOLVED IN THE REPORTED INCIDENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK 20 DEFIBRILLATOR/MONITOR LDD PHYSIO-CONTROL, INC. 20 NA

Patients

Seq Age Sex Outcome Treatment
1 NA