FDA Adverse Event
Malfunction
Summary report: N
LIFEPAK 20 DEFIBRILLATOR/MONITOR
MDR report key: 1003747
·
Received February 22, 2008
Report
- Report Number
- 3015876-2008-00086
- Event Type
- Malfunction
- Date Received
- February 22, 2008
- Date of Event
- January 25, 2008
- Report Date
- January 25, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K012274
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
PHYSIO-CONTROL EVALUATED THE DEVICE AND OBSERVED THAT THE DISPLAY WAS BLANK AND THE SYSTEM WOULD NOT TURN ON. PHYSIO REPLACED THE USER INTERFACE PCB ASSEMBLY AND THEN OBSERVED THAT IT WOULD PROPERLY POWER UP. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE.
Description of Event or Problem · 1
FOUND DURING TESTING. ACCORDING TO THE REPORTER, THE DISPLAY WAS BLANK AND THE DEVICE WOULD NOT TURN ON. THERE WAS NO PATIENT INVOLVED IN THE REPORTED INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 20 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 20 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |