HANDSWITCH PENCIL W/ HOLSTER
Report
- Report Number
- 1717344-2008-00055
- Event Type
- Malfunction
- Date Received
- February 20, 2008
- Date of Event
- January 10, 2008
- Report Date
- January 22, 2008
- Manufacturer
- COVIDIEN (VALLEYLAB)
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE PENCIL WAS TESTED AND MET SPECIFICATION. SPECIFICALLY, TESTING FOUND NO EVIDENCE OF THE BUTTON STICKING ON. THE GENERATOR WAS ALSO RETURNED AND FUNCTIONED TO SPECIFICATION. SOME PHYSICAL DAMAGE WAS FOUND WITH GENERATOR BUT DID NOT EFFECT THE PERFORMANCE OF THE GENERATOR DURING THE EVALUATION. THE GENERATOR AND INCIDENT PENCIL WERE TESTED TOGETHER IN EVERY MODE AND AT ALL POWER LEVELS AND WAS FOUND TO FUNCTION TO SPECIFICATION. THE GENERATOR IS OBSOLETE AND NO LONGER SUPPORTED BY COVIDIEN LP SINCE JULY OF 2005, SO NO REPAIRS WERE DONE. A LETTER WITH THE RESULTS OF OUR EVALUATION, SPECIFYING THE DAMAGE FOUND, WAS SENT TO THE CUSTOMER.
THE REPORT STATED THAT DURING A TAH SURGERY, THE PENCIL HAD BEEN USED AT THE BEGINNING OF THE PROCEDURE. IT WAS SET ASIDE IN A NON-CONDUCTIVE PLASTIC WELL FOR APPROXIMATELY 45 MINUTES. WHEN IT WAS PULLED FROM THE WELL, THE BLADE TIP HAD A QUICK FLAME IGNITION AND THEN WENT OUT. THERE WAS A CUT TONE FROM THE GENERATOR AND THEY REPORTED THAT SWITCH ON THE PENCIL WAS STUCK ON. THEY REMOVED IT FROM THE SURGICAL FIELD AND USED ANOTHER PENCIL TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT. THE BIOMEDICAL ENGINEER AT THE SITE TRIED TO REPLICATE THE INCIDENT, BUT WAS UNABLE. BY THE TIME HE GOT THE PENCIL, THERE WAS NO EVIDENCE OF ANY STUCK BUTTON ON THE PENCIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HANDSWITCH PENCIL W/ HOLSTER | ELECTROSURGICAL ACCESSORY | GEI | COVIDIEN (VALLEYLAB) | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | FORCE28PCH-ELECTROSURGICAL GENERATOR |