SURGAIRTOME TWO HANDPIECE
Report
- Report Number
- 1017294-2008-00134
- Event Type
- Malfunction
- Date Received
- February 20, 2008
- Date of Event
- January 21, 2008
- Report Date
- January 21, 2008
- Manufacturer
- CONMED LINVATEC
- Product Code
- DZA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- DENTAL ASSISTANT
Narratives
INVESTIGATION FINDINGS: CONMED LINVATEC RECEIVED THE DEVICE FORE EVALUATION AND CONFIRMED THE REPORTED PROBLEM. DURING TESTING OF THIS HANDPIECE, IT GOT HOT IN THE COLLET. FURTHER INVESTIGATION FOUND COLLET CORROSION AND A WORN SPINDLE, WHICH PREVENTED THE BUR FROM SECURELY LOCKING INTO THE COLLET. IN ADDITION, THIS UNIT IS OVERDUE FOR PREVENTATIVE MAINTENANCE; LAST SERVICE DATE IDENTIFIED IS FEBRUARY 2006. THE BUR GUARD IN USE AT THE TIME OF THIS EVENT WAS TESTED WITH THIS HANDPIECE AND PERFORMED TO SPECIFICATION. LAST REPAIR DATE FOR THIS BUR GUARD WAS JANUARY 2008, WHICH IS WITHIN THE 6 MONTH PREVENTIVE MAINTENANCE SCHEDULE. AS CERTAIN INTERNAL COMPONENTS (I.E. BEARINGS) ARE KNOWN TO DEGRADE WITH USE, CLEANING, STERILIZATION CYCLES, AND/OR LACK OF PREVENTATIVE MAINTENANCE, CONMED LINVATEC RECOMMENDS THAT THE SURGAIRTOME TWO HANDPIECES BE RETURNED TO THE FACTORY FOR ROUTINE MAINTENANCE EVERY TWELVE (12) MONTHS. FAILURE TO FOLLOW THIS ROUTINE MAINTENANCE SCHEDULE MAY RESULT IN OVERHEATING OR DAMAGE TO THE HANDPIECE AND/OR BUR GUARD AND MAY RESULT IN INJURY TO PATIENTS OR THE MEDICAL PERSONNEL.
THE PHYSICIAN REPORTED THAT DURING USE OF THIS DRILL IN A THIRD MOLAR EXTRACTION, HE FELT HEAT IN THE BODY OF THE HANDPIECE, AND DISCONTINUED USING THE DEVICE. THERE WAS NO REPORT OF SERIOUS INJURY OR SURGICAL DELAY WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SURGAIRTOME TWO HANDPIECE | HIGH SPEED DRILL | DZA | CONMED LINVATEC | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | BUR GUARD| CATALOG #: 1375-12 & LOT #: JAN08S |