FDA Adverse Event Malfunction Summary report: N

SURGAIRTOME TWO HANDPIECE

MDR report key: 1003719 · Received February 20, 2008

Report

Report Number
1017294-2008-00134
Event Type
Malfunction
Date Received
February 20, 2008
Date of Event
January 21, 2008
Report Date
January 21, 2008
Manufacturer
CONMED LINVATEC
Product Code
DZA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTAL ASSISTANT

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS: CONMED LINVATEC RECEIVED THE DEVICE FORE EVALUATION AND CONFIRMED THE REPORTED PROBLEM. DURING TESTING OF THIS HANDPIECE, IT GOT HOT IN THE COLLET. FURTHER INVESTIGATION FOUND COLLET CORROSION AND A WORN SPINDLE, WHICH PREVENTED THE BUR FROM SECURELY LOCKING INTO THE COLLET. IN ADDITION, THIS UNIT IS OVERDUE FOR PREVENTATIVE MAINTENANCE; LAST SERVICE DATE IDENTIFIED IS FEBRUARY 2006. THE BUR GUARD IN USE AT THE TIME OF THIS EVENT WAS TESTED WITH THIS HANDPIECE AND PERFORMED TO SPECIFICATION. LAST REPAIR DATE FOR THIS BUR GUARD WAS JANUARY 2008, WHICH IS WITHIN THE 6 MONTH PREVENTIVE MAINTENANCE SCHEDULE. AS CERTAIN INTERNAL COMPONENTS (I.E. BEARINGS) ARE KNOWN TO DEGRADE WITH USE, CLEANING, STERILIZATION CYCLES, AND/OR LACK OF PREVENTATIVE MAINTENANCE, CONMED LINVATEC RECOMMENDS THAT THE SURGAIRTOME TWO HANDPIECES BE RETURNED TO THE FACTORY FOR ROUTINE MAINTENANCE EVERY TWELVE (12) MONTHS. FAILURE TO FOLLOW THIS ROUTINE MAINTENANCE SCHEDULE MAY RESULT IN OVERHEATING OR DAMAGE TO THE HANDPIECE AND/OR BUR GUARD AND MAY RESULT IN INJURY TO PATIENTS OR THE MEDICAL PERSONNEL.

Description of Event or Problem · 1

THE PHYSICIAN REPORTED THAT DURING USE OF THIS DRILL IN A THIRD MOLAR EXTRACTION, HE FELT HEAT IN THE BODY OF THE HANDPIECE, AND DISCONTINUED USING THE DEVICE. THERE WAS NO REPORT OF SERIOUS INJURY OR SURGICAL DELAY WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SURGAIRTOME TWO HANDPIECE HIGH SPEED DRILL DZA CONMED LINVATEC NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK BUR GUARD| CATALOG #: 1375-12 & LOT #: JAN08S