FDA Adverse Event Injury Summary report: N

STORZ GO KART VIDEO CART MONITORS

MDR report key: 1003447 · Received February 22, 2008

Report

Report Number
MW5005677
Event Type
Injury
Date Received
February 22, 2008
Date of Event
February 22, 2008
Report Date
February 22, 2008
Manufacturer
NATIONAL DISPLAY SYSTEMS
Product Code
KOG
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MO, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

THESE MONITORS ARE SOLD WITH THE STORZ GOKART 9601F VIDEO CARTS THAT WE USE IN SURGERY. THE PROBLEM WE ARE EXPERIENCING IS FREQUENT FAILURES AS A RESULT OF THE MONITOR DESIGN. THE BNC CONNECTORS USED ON THIS MONITOR HAVE NO SUPPORT AROUND THE CENTER CONDUCTOR. AS A DIRECT RESULT OF THIS, THE FEMALE CONNECTION POINTS SPREAD OVER TIME CAUSING INTERFERENCE AND MALFUNCTION. THIS CONNECTOR TYPE DOES NOT WORK WELL IN MOBILE APPLICATIONS SUCH AS THE STORZ GOKART SYSTEMS. WE ARE EXPERIENCING FREQUENT FAILURES CAUSING DELAYS IN SURGICAL CASES. WE BELIEVE THAT THE MFR SHOULD REPLACE THIS CONNECTOR WITH A BNC TYPE CONNECTOR WHOSE CENTER CONDUCTOR IS SUPPORTED TO PREVENT SPREADING AND ULTIMATELY CAUSING A LOOSE CONNECTION. STORZ RADIANCE LCD MONITOR MANUFACTURED BY NATIONAL DISPLAY SYSTEMS MODEL #SC-SX19-A1A11 PART #90X0292-C.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STORZ GO KART VIDEO CART MONITORS VIDEO CART MONITORS KOG NATIONAL DISPLAY SYSTEMS SC-SX19-A1A11

Patients

Seq Age Sex Outcome Treatment
1 YR Disability