ULTRAFLEX COVERED NG ESOPHAGEAL STENT
Report
- Report Number
- 3005099803-2008-00219
- Event Type
- Malfunction
- Date Received
- February 27, 2008
- Date of Event
- November 28, 2007
- Report Date
- February 4, 2008
- Manufacturer
- BOSTON SCIENTIFIC IRELAND, LTD.
- Product Code
- ESW
- PMA / PMN Number
- K032910
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
NOTE: THE EVENT, AS REPORTED, DID NOT REFLECT AN MDR-REPORTABLE SCENARIO; HOWEVER, EVALUATION OF THE RETURNED DEVICE REVEALED AN MDR-REPORTABLE MALFUNCTION. THIS MANUFACTURER REPORT IS SUBMITTED BASED ON THESE RESULTS. A VISUAL EXAMINATION OF THE RETURNED DEVICE REVEALED THAT THE DISTAL SECTION OF THE STENT WAS PARTIALLY DEPLOYED AND HAD FOLDED BACK ON ITSELF. A FUNCTIONAL EVALUATION INDICATED THAT ONCE THE STENT WAS UNFOLDED, IT WAS POSSIBLE TO DEPLOY THE REMAINDER OF THE STENT. A REVIEW OF THE DEVICE HISTORY RECORD WAS PERFORMED FOR THE PERTINENT LOT; NO ANOMALIES WERE NOTED. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO ADDITIONAL COMPLAINTS HAVE BEEN REPORTED FOR THE LOT. THE (B)(6) 2007, 15-MONTH ULTRAFLEX ESOPHAGEAL STENTS PRODUCT FAMILY COMPLAINT TREND REPORT, INCLUSIVE OF ALL FAILURE MODES, WAS REVIEWED; NO UNFAVORABLE TREND WAS NOTED. (B)(4).
ON (B)(6) 2007, IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN ULTRAFLEX COVERED NG ESOPHAGEAL STENT WAS USED DURING A STENT PLACEMENT PROCEDURE ON (B)(6) MALE PATIENT (WEIGHT UNKNOWN) ON (B)(6) 2007. ACCORDING TO THE COMPLAINANT, "THE STENT DIDN'T OPEN PROPERLY.. (T)HE SUTURE DID NOT [BREAK]...[THE PHYSICIAN] DID NOT RELEASE THE STENT...(AND THE STENT WAS REMOVED ALONG) WITH THE INTRODUCER." THE PROCEDURE WAS NOT COMPLETED AS ANOTHER DEVICE WAS NOT AVAILABLE. THE PATIENT'S CONDITION WAS REPORTED AS "STABLE".
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ULTRAFLEX COVERED NG ESOPHAGEAL STENT | ESW | BOSTON SCIENTIFIC IRELAND, LTD. | M00514200 | 9344125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |