VITROS 250 CHEMISTRY SYSTEM
Report
- Report Number
- 1319681-2008-00056
- Event Type
- Malfunction
- Date Received
- February 27, 2008
- Date of Event
- January 26, 2008
- Report Date
- January 30, 2008
- Manufacturer
- ORTHO-CLINICAL DIAGNOSTICS
- Product Code
- JJE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE INVESTIGATION BY OCD DETERMINED THAT RESULTS FROM ONE PATIENT SAMPLE WAS ASSOCIATED WITH AN INCORRECT SAMPLE IDENTIFICATION NUMBER. OCD HAS EVALUATED INSTRUMENT DATALOGGERS AND DETERMINED THAT THE OPERATOR HAD MANUALLY ENTERED INFORMATION FOR THE SAMPLE ID IN QUESTION. FURTHER EVALUATION OF THE DATALOGGERS ALONG WITH THE OPERATORS STATEMENTS INDICATE THAT THE ANALYZER PERFORMED AS EXPECTED. IT APPEARS FROM THE DATALOGGER OUTPUT THAT THE ISSUE MAY BE RELATED TO THE MANUAL PROGRAMING OF THE SAMPLE ID BUT BECAUSE OF PATIENT CONFIDENTIALITY THIS IS NOT ABLE TO BE CONFIRMED. SAMPLE TRAY POSITIONING ADJUSTMENTS AND SAMPLE BAR CODE READINGS WERE CHECKED AND DETERMINED TO BE WORKING AS EXPECTED WHEN EVALUATED AFTER THE EVENT. THE OPERATOR SUBSEQUENTLY PLACED THE SAMPLES BACK ON THE INSTRUMENT AND THEY RAN WITH RESULTS AS EXPECTED. NO INCORRECT PATIENT RESULTS WERE REPORTED AND THERE WAS NO ALLEGATION OF HARM AS A RESULT OF THIS EVENT. THE MOST LIKELY CAUSE OF THE EVENT APPEARS TO BE OPERATOR ERROR, BUT BECAUSE OF THE INABILITY TO EVALUATE SAMPLE ID INFORMATION, THIS CANNOT BE CONFIRMED, THEREFORE NO CONCLUSIVE ROOT CAUSE CAN BE DETERMINED.
THE OPERATOR OF A VITROS 250 CHEMISTRY SYSTEM OBSERVED PATIENT RESULTS ASSOCIATED WITH INCORRECT SAMPLE IDENTIFICATION NUMBER. THE LABORATORY DID NOT REPORT ANY RESULTS FOR THE SAMPLE DETERMINED TO HAVE AN INCORRECT SAMPLE IDENTIFICATION NUMBER, AND THERE WAS NO ALLEGATION OF HARM. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VITROS 250 CHEMISTRY SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | ORTHO-CLINICAL DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |