FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PLUS PROFILE

MDR report key: 10032976 · Received May 7, 2020

Report

Report Number
1645337-2020-06112
Event Type
Injury
Date Received
May 7, 2020
Report Date
April 15, 2020
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001485
PMA / PMN Number
P990075
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

ON (B)(6) 2020, MENTOR FAILURE ANALYSIS LAB COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION REPORTED, THE PATIENT DEVELOPED CAPSULAR CONTRACTURE. DURING THE VISUAL EVALUATION, AN ANOMALY WAS OBSERVED ON THE ANTERIOR VIEW OF THE DEVICE, CONSISTENT WITH A CREASE/FOLD. MENTOR CONCLUDED THAT THE CAPSULAR CONTRACTURE IN THE PATIENT´S BREAST WAS THE RESULT OF THE BODY´S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. A SECOND PRODUCT WAS RECEIVED (PRODUCT CODE 3502550 AND LOT NUMBER 6901156). NO ADVERSE EVENTS WERE REPORTED FOR THIS CONCOMITANT (CONTRALATERAL) DEVICE, THEREFORE NO FURTHER ANALYSIS IS REQUIRED. AS PART OF OUR QUALITY PROCESS, THE MANUFACTURING RECORDS OF LOT 6901156 WERE REVIEWED AND THE MANUFACTURING STANDARDS WERE MET PRIOR TO THE RELEASE OF THIS LOT. EACH DEVICE IS VISUALLY INSPECTED DURING MANUFACTURING TO ENSURE THE DEVICE MEETS THE REQUIRED SPECIFICATIONS PRIOR TO SHIPMENT. CAPSULAR CONTRACTURE COMPLAINT INFORMATION IS CONSISTENTLY ANALYZED AND MONITORED BY MENTOR QUALITY ASSURANCE TO DETERMINE WHEN FURTHER ACTION IS NECESSARY. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE IV MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT A CAUCASIAN FEMALE PATIENT WHO UNDERWENT A BREAST AUGMENTATION REVISION WITH 550CC SMOOTH MENTOR SMOOTH ROUND MODERATE PLUS PROFILE SALINE BREAST IMPLANT HAS EXPERIENCED POSTOPERATIVE RIGHT-SIDED GRADE IV CAPSULAR CONTRACTURE. AS A RESULT, THE PATIENT UNDERWENT EXPLANTATION WITHOUT REPLACEMENT ON (B)(6) 2019.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
498416 MENTOR SMOOTH ROUND MODERATE PLUS PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 3502550 6901156 00081317001485

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention