GEMINI 11MM BASKET
Report
- Report Number
- 3005099803-2008-00189
- Event Type
- Injury
- Date Received
- February 21, 2008
- Date of Event
- January 29, 2008
- Report Date
- January 30, 2008
- Manufacturer
- BOSTON SCIENTIFIC CORPORATION
- Product Code
- FFL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE COMPLAINANT INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS IS NOT AVAILABLE, AND WE ARE NOT ABLE TO DETERMINE THE RELATIONSHIP BETWEEN THIS DEVICE AND THE CAUSE FOR THIS EVENT. A REVIEW OF THE DEVICE HISTORY (DHR) WAS PERFORMED ON THE PERTINENT LOT; NO ANOMALIES WERE NOTED. A LOT HISTORY SEARCH WAS PERFORMED AND FOUND NO OTHER COMPLAINTS HAVE BEEN REPORTED FROM THIS LOT. THE DECEMBER 2007 15-MONTH 11MM GEMINI COMPLAINT TREND REPORT, INCLUSIVE OF ALL FAILURE MODES, WAS REVIEWED, NO ADVERSE TREND WAS NOTED. ATTEMPTING TO REMOVE A STONE THAT IS TOO LARGE IS CONTRAINDICATED IN THE DFU (DIRECTIONS FOR USE).
IN 2008, IT WAS REPORTED TO BOSTON SCIENTIFIC CORP THAT A GEMINI RETRIEVAL BASKET WAS USED IN A STONE REMOVAL PROCEDURE ON A MALE PT (WEIGHT UNK) THAT OCCURRED ON THE DAY BEFORE. DURING THE PROCEDURE, THE PHYSICIAN ATTEMPTED TO EXTRACT A LARGE STONE FROM THE PT BUT THE BASKET BECAME STUCK. CONSEQUENTLY, THE PHYSICIAN OPTED TO CUT THE HANDLE AND LEFT THE BASKET INSIDE THE PT. THE PT WAS ADMITTED TO ANOTHER HOSP FOR FURTHER TREATMENT. A STENT WAS PLACED AND THE PT WAS KEPT OVERNIGHT FOR OBSERVATION. ACCORDING TO THE COMPLAINANT, THE BASKET OF THE GEMINI RETRIEVAL BASKET DEVICE WAS RETRIEVED (UNK METHOD) AND THE LARGE STONE WAS REMOVED BY A LASER PROCEDURE. A USER FACILITY MEDWATCH WAS RECEIVED ON 02/15/2008.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GEMINI 11MM BASKET | FFL | FFL | BOSTON SCIENTIFIC CORPORATION | M0063301040 | 11327380 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Hospitalization| R |