ADVANTA BED
Report
- Report Number
- 1824206-2008-00004
- Event Type
- Injury
- Date Received
- January 25, 2008
- Date of Event
- June 15, 2007
- Report Date
- January 16, 2008
- Manufacturer
- HILL-ROM RITTER
- Product Code
- FNL
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
ON JAN 16, 2008 HILL-ROM RECEIVED A COPY OF MEDWATCH, INDICATING THAT A PT FALL OCCURRED RESULTING IN A BROKEN HIP. HILL-ROM WAS ORIGINALLY NOTIFIED OF A PT FALL BY THE FACILITY IN 2007 BUT THE FACILITY WOULD NOT RELEASE ANY DETAILS OF THE INCIDENT AT THAT TIME. ALSO THE FACILITY COULD NOT PROVIDE A MODEL OR SERIAL NUMBER OF THE BED. WHEN THE PT EXITED THE BED A PRIORITY NURSE CALL ALARM WAS NOT RECEIVED AT THE NURSE'S STATION. THE HILL-ROM TECHNICIAN FOUND NO MALFUNCTION ON THE HILL-ROM BED AND THAT THE BED EXIT WORKED AS DESIGNED. HE DID FIND THAT THE WESCOM NURSE CALL SYSTEM WAS NOT COMPATIBLE WITH THE HILL-ROM BED ON PLACING A PRIORITY NURSE CALL. THE FACILITY HAS PURCHASED SIDECOME CABLES FROM CORBELL THAT WILL MAKE THE NURSE CALL SYSTEM COMPATIBLE WITH THE HILL-ROM BEDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADVANTA BED | AC POWERED HOSPITAL BED | FNL | HILL-ROM RITTER | P1600 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 89 YR | Other |