PRISM 2000 GAMMA CAMERA
Report
- Report Number
- 1525965-2008-00006
- Event Type
- Other
- Date Received
- February 8, 2008
- Report Date
- February 8, 2008
- Manufacturer
- PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
- Product Code
- IYX
- PMA / PMN Number
- K904050
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
INITIAL INVESTIGATION DETERMINED THE FOLLOWING: THE COLLIMATOR ID BOARD ON HEAD 2 WAS SHIMMED OUT. AS A RESULT, THE DETECTOR FALSELY SENSED THE COLLIMATOR WAS LATCHED. THE BLOCK ON THE ID BOARD WAS DAMAGED. IT WAS DETERMINED THAT THE SYSTEM WAS UNDER THIRD PARTY SERVICE. THE LAST PREVENTIVE MAINTENANCE WAS CONDUCTED IN 2006. ENGINEERING REVIEW OF THE PHOTS TAKEN AT THE SITE DETERMINED THAT FURTHER INVESTIGATION AT THE SITE WAS NECESSARY. DURING A SITE VISIT, A PHILIPS TECHNICAL SUPPORT ENGINEER OBSERVED THE FOLLOWING: THE GANTRY COVERS WERE MISSING SEVERAL CLIPS. THIS PREVENTS A SECURE ATTACHMENT OF THE FRONT AND REAR COVER. THERE IS A COLLECTION OF METAL SHAVINGS ON THE DETECTOR. THIS WAS MOST LIKELY THE RESULT OF THE RUBBER CUSHION ON THE DETECTOR BEING VERY WORN. WHEN THE DETECTOR IS ROTATED, THE SHAVINGS COULD FALL IN THE PT'S EYES. THE COLLIMATOR SERVER CART HAD BEEN MODIFIED BY REMOVING THE TOP SAFETY CATCHES (INTENDED TO PREVENT THE COLLIMATOR FROM FALLING DURING EXCHANGE). THE WRAP AROUND THE GANTRY POWER CABLES WAS SPLIT. ALL THE CABLES INSIDE WERE IN GOOD CONDITION AND NONE OF THE INSULATION WAS DAMAGED. THE MOST PROBABLE ROOT CAUSE OF THESE ISSUES IS IMPROPER MAINTENANCE PERFORMED BY THE CUSTOMER. THERE WAS NO SERIOUS INJURY TO PT OR OPERATOR.
TECHNICIAN WAS LOADING THE MEGAP-PAR (MEDIUM ENERGY ALL PURPOSE-PARALLEL) COLLIMATOR ON TO DETECTOR HEAD 2 FOR A NEW PT SCAN. MEGAP-PAR COLLIMATOR SEPARATED FROM THE DETECTOR HEAD SHORTLY AFTER LOADING. THE COLLIMATOR WAS HELD TO THE HEAD BY SAFETY LATCH. NO INJURY TO PT OR OPERATOR OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PRISM 2000 GAMMA CAMERA | GAMMA CAMERA | IYX | PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. | 4535 665 12911 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |