FDA Adverse Event Other Summary report: N

PRISM 2000 GAMMA CAMERA

MDR report key: 1002924 · Received February 8, 2008

Report

Report Number
1525965-2008-00006
Event Type
Other
Date Received
February 8, 2008
Report Date
February 8, 2008
Manufacturer
PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC.
Product Code
IYX
PMA / PMN Number
K904050
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INITIAL INVESTIGATION DETERMINED THE FOLLOWING: THE COLLIMATOR ID BOARD ON HEAD 2 WAS SHIMMED OUT. AS A RESULT, THE DETECTOR FALSELY SENSED THE COLLIMATOR WAS LATCHED. THE BLOCK ON THE ID BOARD WAS DAMAGED. IT WAS DETERMINED THAT THE SYSTEM WAS UNDER THIRD PARTY SERVICE. THE LAST PREVENTIVE MAINTENANCE WAS CONDUCTED IN 2006. ENGINEERING REVIEW OF THE PHOTS TAKEN AT THE SITE DETERMINED THAT FURTHER INVESTIGATION AT THE SITE WAS NECESSARY. DURING A SITE VISIT, A PHILIPS TECHNICAL SUPPORT ENGINEER OBSERVED THE FOLLOWING: THE GANTRY COVERS WERE MISSING SEVERAL CLIPS. THIS PREVENTS A SECURE ATTACHMENT OF THE FRONT AND REAR COVER. THERE IS A COLLECTION OF METAL SHAVINGS ON THE DETECTOR. THIS WAS MOST LIKELY THE RESULT OF THE RUBBER CUSHION ON THE DETECTOR BEING VERY WORN. WHEN THE DETECTOR IS ROTATED, THE SHAVINGS COULD FALL IN THE PT'S EYES. THE COLLIMATOR SERVER CART HAD BEEN MODIFIED BY REMOVING THE TOP SAFETY CATCHES (INTENDED TO PREVENT THE COLLIMATOR FROM FALLING DURING EXCHANGE). THE WRAP AROUND THE GANTRY POWER CABLES WAS SPLIT. ALL THE CABLES INSIDE WERE IN GOOD CONDITION AND NONE OF THE INSULATION WAS DAMAGED. THE MOST PROBABLE ROOT CAUSE OF THESE ISSUES IS IMPROPER MAINTENANCE PERFORMED BY THE CUSTOMER. THERE WAS NO SERIOUS INJURY TO PT OR OPERATOR.

Description of Event or Problem · 1

TECHNICIAN WAS LOADING THE MEGAP-PAR (MEDIUM ENERGY ALL PURPOSE-PARALLEL) COLLIMATOR ON TO DETECTOR HEAD 2 FOR A NEW PT SCAN. MEGAP-PAR COLLIMATOR SEPARATED FROM THE DETECTOR HEAD SHORTLY AFTER LOADING. THE COLLIMATOR WAS HELD TO THE HEAD BY SAFETY LATCH. NO INJURY TO PT OR OPERATOR OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRISM 2000 GAMMA CAMERA GAMMA CAMERA IYX PHILIPS MEDICAL SYSTEMS (CLEVELAND), INC. 4535 665 12911

Patients

Seq Age Sex Outcome Treatment
1 UNK