FDA Adverse Event Other Summary report: N

DA VINCI SURGICAL SYSTEM

MDR report key: 1002915 · Received February 11, 2008

Report

Report Number
2955842-2008-00055
Event Type
Other
Date Received
February 11, 2008
Date of Event
January 17, 2008
Report Date
February 8, 2008
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ON SITE INVESTIGATION AND EVALUATION WAS CONDUCTED BY FIELD SERVICE ENGINEERING. SYSTEM VERIFICATION TESTS WERE PERFORMED AND THE CUSTOMER REPORTED FAILURE MODE COULD NOT BE DUPLICATED. THE RIGHT PT SIDE MANIPULATOR (PSM) WAS REPLACED AS A PRECAUTION. THE PT SIDE MANIPULATOR IS AN INSTRUMENT ARM LOCATED ON THE PT SIDE CART WHICH PROVIDES THE STERILE INTERFACE FOR THE ENDOWRIST INSTRUMENTS. THE PSM WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. MECHANICAL AND ELECTRICAL PERFORMANCE TESTS WERE CONDUCTED AND THE CUSTOMER REPORTED FAILURE MODE COULD NOT BE REPRODUCED. AS OF 2008, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY SURGICAL PROCEDURE, THE RIGHT SYSTEM ARM WAS NOT ALLOWING THE INSTRUMENT JAWS TO OPEN AND CLOSE PROPERLY, AND THE SURGEON FOUND IT DIFFICULT TO ARTICULATE THE JAWS DURING WRIST MOVEMENT. THE INSTRUMENT STERILE ADAPTER WAS REPLACED, HOWEVER, THESE ISSUES CONTINUED TO RECUR. NO PATIENT HARM WAS REPORTED. THE PROCEDURE WAS CONVERTED TO TRADITIONAL OPEN SURGICAL TECHNIQUES TO FINISH THE PLANNED SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DA VINCI SURGICAL SYSTEM ENDOSCOPIC INSTRUMENT CONTROL SYSTEM NAY INTUITIVE SURGICAL, INC. IS12000 A4.3P8

Patients

Seq Age Sex Outcome Treatment
1 Other DA VINCI SURGICAL INSTRUMENTS| AND ACCESSORIES