FDA Adverse Event Injury Summary report: N

PICCOLO COMPOSITE PROXIMAL FEMUR NAIL

MDR report key: 10028610 · Received May 6, 2020

Report

Report Number
9615128-2020-00005
Event Type
Injury
Date Received
May 6, 2020
Date of Event
March 19, 2020
Report Date
June 5, 2020
Manufacturer
CARBOFIX ORTHOPEDICS LTD.
Product Code
HSB
UDI-DI
07290109336534
PMA / PMN Number
K153536
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE INVOLVED ITEM WAS RECEIVED AND UNDERWENT AN EXAMINATION BY THE COMPANY. EXAMINATION OF THE NAIL, INCLUDING MICROSCOPIC VISUAL INSPECTION, INDICATED IT HAD A FATIGUE FRACTURE. EXAMINATION OF THE PRODUCTION RECORDS OF THE INVOLVED DEVICE INDICATED THE IMPLANT WAS MANUFACTURED ACCORDING TO SPECIFICATION. BREAKAGE OF PROXIMAL FEMUR NAILS IS KNOWN AND DOCUMENTED IN THE LITERATURE, WITH REPORTED RATES THAT RANGE FROM 0.2% TO 5.6%(1). THE RISK OF IMPLANT FATIGUE FAILURE INCREASES IN THE CASE OF PATHOLOGICAL FRACTURES, LIKE IN THE REPORTED CASE. THE REPORTED CASE INVOLVED A (B)(6) YEAR OLD PATIENT WITH A TUMOR. IMPORTANTLY, NAIL WITHHELD FOR MORE THAN 9 MONTHS WHICH IS CONSIDERED A LONG PERIOD IN CASES WITH COMPROMISED BONE HEALING. (1) JOHNSON NA ET AL., RISK FACTORS FOR INTRAMEDULLARY NAIL BREAKAGE IN PROXIMAL FEMORAL FRACTURES: A 10-YEAR RETROSPECTIVE REVIEW. ANN R COLL SURG ENGL. 2017; 99(2): 145-150.

Description of Event or Problem · 1

A PROXIMAL FEMUR NAIL WAS IMPLANTED TO TREAT A PATHOLOGIC FRACTURE (ONCOLOGICAL PATIENT) IN (B)(6). MORE THAN 9 MONTHS POST-OPERATION, NAIL BROKE AT THE SUB-TROCHANTERIC AREA AND WAS REPLACED WITH A METAL NAIL IN A REVISION SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
491557 PICCOLO COMPOSITE PROXIMAL FEMUR NAIL INTRAMEDULLARY FIXATION ROD SYSTEM HSB CARBOFIX ORTHOPEDICS LTD. Proximal Femur Nail PPF01196 07290109336534

Patients

Seq Age Sex Outcome Treatment
1 72 YR Required Intervention