FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 1002760
·
Received February 26, 2008
Report
- Report Number
- 1720753-2008-16392
- Event Type
- Malfunction
- Date Received
- February 26, 2008
- Date of Event
- February 15, 2008
- Report Date
- February 26, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS, INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. RESEATED BOARDS AND PERFORMED AN MA NULL CALIBRATION AND A FILAMENT CALIBRATION. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM PRESENTED A MA SENSOR FAIL INDICATION. THERE WAS NO REPORT OF PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS, INC. | 9800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |