FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 1002742 · Received February 26, 2008

Report

Report Number
1720753-2008-16410
Event Type
Malfunction
Date Received
February 26, 2008
Date of Event
February 19, 2008
Report Date
February 26, 2008
Manufacturer
GE OEC MEDICAL SYSTEMS INC
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE MALFUNCTION WAS VERIFIED. REPLACED THE DISPLAY ADAPTER, IMAGE PROCESSOR, GENERATOR INTERFACE PCB, FLUORO FUNCTION BOARD PCB AND THE SNUBBER PCB. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PLACED BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE LEFT MONITOR ON THE 9800 SYSTEM WENT BLACK DURING THE CASE. THERE WAS NO REPORT OF PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS INC 9800 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK