FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE-PIECE FOLDABLE INTRAOCUL

MDR report key: 1002710 · Received February 26, 2008

Report

Report Number
2023826-2008-00273
Event Type
Malfunction
Date Received
February 26, 2008
Date of Event
January 22, 2008
Report Date
January 29, 2008
Manufacturer
STAAR SURGICAL
Product Code
HQL
PMA / PMN Number
P880091
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL RESULTS: A VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWS THE LENS IS TORN IN HALF FROM ONE HAPTIC TO THE OTHER HAPTIC. THERE IS CLEAR SURGICAL RESIDUE ON THE LENS. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVALUATION. CONCLUSIONS: AN INVESTIGATION WAS OPENED TO EVALUATE A COMPLAINT TREND ASSOCIATED WITH LENS TEARS THAT WAS ORIGINALLY IDENTIFIED IN JUNE 2005. POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND POSSIBLE HANDLING ERRORS BY THE CUSTOMER. TO ADDRESS DELIVERY SYSTEM ISSUES, ALL STAGES IN THE MANUFACTURING OF THE INJECTORS AND CARTRIDGES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. TO ADDRESS HANDLING ERRORS, ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THAT THE SURGEON WAS INSERTING A ONE PIECE SILICONE LENS MODEL AA4204VF AND THE LENS TORE. THE LENS WAS REMOVED WITH NO INCISION ENLARGEMENT OR SUTURE REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE-PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL AA4204VF NA

Patients

Seq Age Sex Outcome Treatment
1 VISCOELASTIC: STAARVISC II LOT NUMBER: UNK| CARTRIDGE MODEL: MTC-60CFP| INJECTOR MODEL: MSI-PR